After that the several different measurement values can be indicated using the UP or DOWN keys.
Keep the DOWN or UP key pressed for > 3 sec to exit the MIN / MAX mode.
Maximal
V
: 323.8 m
0
5.3.4.
Reset of the consumption counter or the MIN / MAX value
Keep both the UP and DOWN key pressed for > 3 sec to enter the menu for resetting the consumption coun-
ter or the MIN / MAX value. Select the desired menu item by pressing the UP or DOWN key briefly.
Rel : 1
V
: 45.0 m
0
Maximal
V
: 323.8 m
0
To confirm the selected choice of the menu keep the DOWN or UP key pressed for > 3 sec.
Select menu item "NO" or "EXIT" to cancel without resetting.
Rel : 1
V
: 45.0 m
0
> Clear Consumption
Clear Min/Max
Exit
5.3.5.
Maximum consumption counter
The maximum consumption counter readout on the display is 999,999,999.0 m
shows "LCD maximum". The internal memory continues counting. The maximum counter reading is 3.4 * 10
m
. It is possible to read-out the counter reading with the configuration software.
3
6.
ERROR MESSAGES
In case the flow meter is equipped with the optional display, the following error message can be indicated.
ERROR 01: sensing probe is not detected
Cause:
the sensing probe is not connect or is defect
Effect:
the display indicates "0" for all measurand. The analogue output defaults to the lowest
configured value.
Action:
check the head of the sensing probe for visual damage.
check the sensor cable from the sensing probe to the electronics of the signal conditioning.
ERROR 02: the EEprom is defect
Cause:
the EEPROM for the storing of the consumption counter and MIN /MAX value is defect.
Effect:
the consumption counter and MIN / MAX values are no longer available. All measure-
ment values though are still indicated. The analogue, relay and pulse output are still
functional.
Action:
return the flow meter to the manufacturer.
/h
3
/h
3
Press pushbutton UP and DOWN at the same time for >3s
/h
3
/h
3
18
or 99,999,999.0 ft
. Then it
3
3
38