Factory Setting Of Outputs - E+E Elektronik EE771 Series Manual

Flow meter for compressed air and gases
Table of Contents

Advertisement

General
Supply voltage
Current consumption
Temperature range
Nominal pressure
Humidity
Medium
Connection
Electromagnetic compatibility
Material
Housing protection class
1) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was
culated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
9.1.

Factory setting of outputs

SI-Unit
Analogue output
[0...10 V / 0(4)...20 mA]
standardized
volumetric flow
mass flow
standardized flow
temperature
Switching output
standardized
volumetric flow
mass flow
standardized flow
temperature
Pulse output
housing
probe
sensor head
measurement ball valve
from
low (L1)
DN15:
0
60
DN20:
0
110
DN25:
0
175
DN32:
0
285
DN40:
0
450
DN50:
0
700
DN15:
0
75
DN20:
0
140
DN25:
0
220
DN32
0
360
DN40:
0
570
DN50:
0
890
0
100
≤DN50
-20
80
all Ø
[switching point/hysteresis]
DN15
50/5
DN20
90/9
DN25:
150/15
DN32:
230/23
DN40:
360/36
DN50:
560/56
DN15:
60/6
DN20:
110/11
DN25:
200/20
DN32:
290/29
DN40:
460/46
DN50:
700/70
80/8
≤DN50
30/3
all Ø
pulse-value = 1 m
3
18 - 30 V AC/DC
max. 200 mA (with display)
ambient temperature:
medium temperature:
storage temperature:
up to 16 bar
(232 psi)
no condensation
compressed air or none corrosive gases
cable gland M16x1,5 (optional connector M12x1 8pol.)
EN61326-1
Industrial Environment
metal (AlSi3Cu)
stainless steel
stainless steel / glass
brass
IP65 / Nema 4
to
high (H1)
120
220
350
570
900
1400
150
280
440
720
1140
1780
200
80
100/10
180/18
300/30
460/46
720/72
1120/112
120/12
220/22
400/40
580/58
920/92
1400/140
180/18
70/7
pulse-duration = 0,1 sec.
21
-20...60 °C
(-4...140 °F)
-20...80 °C
(-4...176 °F)
-20...60 °C
(-4...140 °F)
EN61326-2-3
unit
Nm
/h
3
Nm
/h
3
Nm
/h
3
Nm
/h
3
Nm
/h
3
Nm
/h
3
kg/h
kg/h
kg/h
kg/h
kg/h
kg/h
Nm/s
°C
Nm
/h
3
Nm
/h
3
Nm
/h
3
Nm
/h
3
Nm
/h
3
Nm
/h
3
kg/h
kg/h
kg/h
kg/h
kg/h
kg/h
Nm/s
°C

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ee771-bEe771-aEe771-c

Table of Contents