ATEX Instruction Manual
ETC01035
January 2004
Depending on which system is used, diffe-
rent parameters apply:
Application
Zone 1
for measuring
flammable gases
Zone 1
for measuring
non-flammable gases
Application
Zone 2
for measuring
non-flammable gases
Minimum overpressure in all enclosures:
1 mbar against external atmosphere and
•
50 Pa against containment system when measuring flammable gases
•
Emerson Process Management Manufacturing GmbH & Co. OHG
Mode of Operation
Prepurge Phase
Duration:
Purge gas consumption:
maximum overpressure:
Measuring Mode
Internal overpressure:
Flow rate (leakage rate):
Prepurge Phase
Duration:
Purge gas consumption:
maximum overpressure:
Measuring Mode
Internal overpressure:
Flow rate (leakage rate):
Mode of Operation
Prepurge Phase
Duration:
Purge gas consumption:
maximum overpressure:
Measuring Mode
Internal overpressure:
Flow rate (leakage rate):
MLT 2 / BINOS
Single Enclosure
(volume approx. 56 l)
10 min
1,75 Nm³/h
95 mbar
90mbar
<= 3 l/min
4 min
6,8 Nm³/h
25 mbar
14 mbar
<= 3 l/min
Single Enclosure
(volume approx. 56 l)
5 min
8 Nm³/h
25 mbar
1 mbar
<= 22Nm³/h
®
100 F
Dual Enclosure (volume
approx. 112 l)
14 min
1,92 Nm³/h
95 mbar
90 mbar
<= 4,5 l/min
6 min
9,2 Nm³/h
25 mbar
14 mbar
<= 4,5 l/min
Dual Enclosure (volume
approx. 112 l)
17 min
7 Nm³/h
25 mbar
1 mbar
<= 22 Nm³/h
1 - 5