Section 4 Maintenance; Maintenance Interval; Gasanalyzer; Pressurization Systems And Other Additional Equipment - Emerson Rosemount BINOS 100 F Instruction Manual

Addendum for pressurized analyzers intended to be used in hazardous areas: zone 1: all versions, zone 2: non-flammable gases only
Hide thumbs Also See for Rosemount BINOS 100 F:
Table of Contents

Advertisement

ATEX Instruction Manual
ETC01035
August 2005
After maintenance or replacement of parts concerning explosion protection an
authority on explosion protection has to verify that the analyzer still meets the
requirements for explosion protection before it is switched on again.
If parts essential for explosion protection are repaired they have to be routine
tested!
The authority has to issue a certificate for this and/or attach a test label to the
equipment before startup after maintenance or replacement of parts.
4-1

Maintenance Interval

To ensure the performance and safety of the
equipment it has to be checked on a regular
basis, at least once a year. Special care has
to be taken for the EEx p saftey device and
parts ensuring explosion protection (e.g.
gaskets).
4-2

Gasanalyzer

Refer to the associated manual for detailed
information on maintenance, replacement of
parts and how to carry out a containment
system leak test.
4-3
Pressurization Systems and
Other Additional Equipment
Refer to the associated manuals for detailed
information
on
replacement of parts.
Emerson Process Management Manufacturing GmbH & Co. OHG
SECTION 4
Maintenance
maintenance
and
MLT 2 / BINOS
It is the operators/owners
responsibility to extend the
maintenance interval with
respect
to
influences of gases or
environment on materials in
contact with the sample gas
or
ensuring
protection (e.g. gaskets).
Defective parts relevant for
the explosion protection
must only be replaced by ori-
ginal Emerson Process Ma-
nagement spare parts!
Unauthorized parts and
procedures can affect the
product's performance, place
the safe operation of your
process at risk and affect the
explosion protection!
®