Equivalent Safety Devices - Emerson Rosemount BINOS 100 F Instruction Manual

Addendum for pressurized analyzers intended to be used in hazardous areas: zone 1: all versions, zone 2: non-flammable gases only
Hide thumbs Also See for Rosemount BINOS 100 F:
Table of Contents

Advertisement

MLT 2 / BINOS
1-3-7

Equivalent Safety Devices

It is recommended to order and use the pressurization systems provided and
specified by Emerson Process Management.
It is the responsibility of the operator/owner to choose and install a pressurization
system other than recommended by Emerson Process Management!
In no case Emerson Process Management is held responsible for any damage,
personal injury etc. resulting from choosing different pressurization systems!
In principle it is possible to use other
equivalent pressurization systems as
described in this manual, as long as they
meet the following requirements:
• Device is ATEX approved
• Adjustable pre-purge phase
• Minimum purge flow rate :
7 Nm
• Limited purge gas pressure during the
pre-purging phase: < 25 mbar
• Input pressure for purge gas supply:
2 ... 4 bar
• Relay contacts for all-pole disconnection
of the analyzer from power supply
• Relay contacts for disconnecting additi-
onal external equipment
• Connection facility for bypass key switch,
providing intrinsically safe signal for
scanning the key switch
1 - 8
®
100 F
1-3-7
Equivalent Safety Devices
3
/h resp. 8 Nm
3
/h
ATEX Instruction Manual
• For units connected to flammable
gases, only:
Connection facility for differential
pressure switch, providing intrinsically
safe signal for scanning the switch
Emerson Process Management Manufacturing GmbH & Co. OHG
ETC01035
January 2004