Instruction Manual ETC01035 08/2005 Instruction Manual BINOS 100 F & MLT 2 Addendum for Pressurized Analyzers intended to be used in Hazardous Areas Zone 1: All Versions Zone 2: Non-Flammable Gases Only Edition 08/2005 www.EmersonProcess.com...
Page 2
• Read all instructions prior to installing, operating, and servicing the product. • If you do not understand any of the instructions, contact your Emerson Process Management (Rosemount Analytical) representative for clarification. • Follow all warnings, cautions, and instructions marked on and supplied with the product.
Some technical specifications in this manual may be different to those in the associated manuals for the analyzers and additional equipment. In this case the technical specifications listed in this manual are valid only. Emerson Process Management Manufacturing GmbH & Co. OHG P - 1...
Products described in this manual are not designed for use in any life support and/or safety equipment where failure to perform can reasonably be expected to result in personal injury or death. Emerson Process Management Manufacturing GmbH & Co. OHG P - 3...
Page 6
To avoid loss of life, personal injury and damage to this equipment and on-site property, do not operate or service this instrument before reading and understanding this instruction manual and receiving appropriate training. SAVE THESE INSTRUCTIONS FOR FUTURE USE. P - 4 Emerson Process Management Manufacturing GmbH & Co. OHG...
(gas) to burn. rich in fuel (deficient in oxygen) to burn. Emerson Process Management Manufacturing GmbH & Co. OHG P - 5...
Page 8
> 1 atmospheric. mbar compared to atmospheric. Flammable Gas(es) Gases and gas mixtures are assigned to be flammable if they might become ignitable when in a mixture with air. P - 6 Emerson Process Management Manufacturing GmbH & Co. OHG...
Page 9
ATEX Instruction Manual ETC01035 ® MLT 2 / BINOS 100 F August 2005 Emerson Process Management Manufacturing GmbH & Co. OHG P - 7...
(EEx p) using simplified Applicable Standards: purge with either leakage EN 50014:1997 + amendments 1 & 2, compensation mode or EN 50016:1995, continuous flow EN 50020:1994 Temperature class: T4 Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 3...
(e.g. containment system. prepurge phase, purge gas flow, internal overpressure, and more) and therefore ensures safe operation. A separate instruc- tion manual provides more detailled infor- mation. 1 - 4 Emerson Process Management Manufacturing GmbH & Co. OHG...
Page 16
Flow rate (leakage rate): <= 22Nm³/h <= 22 Nm³/h Minimum overpressure in all enclosures: 1 mbar against external atmosphere and • 50 Pa against containment system when measuring flammable gases • Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 5...
Medium has to be dry and free from dust, oil, corrosive or aggressive components! Input pressure at inlet of EEx p device: • 2.000 to 4.000 hPa (2 to 4 bar) 1 - 6 Emerson Process Management Manufacturing GmbH & Co. OHG...
The limit of 50 Pa is the least permissible value given by the standards, higher values are acceptable, too. Flame arrestors are required for gas inlets and outlets. Electrical and mechanical connections are shown in section 2 „Installation“. Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 7...
It is the responsibility of the operator/owner to choose and install a pressurization system other than recommended by Emerson Process Management! In no case Emerson Process Management is held responsible for any damage, personal injury etc. resulting from choosing different pressurization systems! •...
The maximum permissible gas • pressure depends on the type of connected gas. See section 1-3-5 for detailed information. Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 9...
Page 21
ATEX Instruction Manual ETC01035 ® MLT 2 / BINOS 100 F January 2004 1 - 10 Emerson Process Management Manufacturing GmbH & Co. OHG...
Do not use the attached EEx p unit or the external valve as a handle for transportation or lifting the instrument! Failure to follow may loosen or damage the devices and impair safety of the whole instrument! Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 1...
MLT 2 / BINOS 100 F August 2005 2-3 Dimensional Drawings Dimensional Drawings 2-3-1 Single Enclosure Analyzer Standard version w/o purging system Fig. 2-1a Single enclosure analyzer (approx. dimensions in mm) 2 - 2 Emerson Process Management Manufacturing GmbH & Co. OHG...
Page 24
ATEX Instruction Manual ETC01035 ® MLT 2 / BINOS 100 F August 2005 2-3-1 Single Enclosure Analyzer Fig. 2-1b Single enclosure analyzer (approx. dimensions in mm) Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 3...
August 2005 2-3-2 Dual Enclosure Analyzer 2-3-2 Dual Enclosure Analyzer Standard version w/o purging system with category 3 purging system Fig. 2-2a Dual enclosure analyzer (approx. dimensions in mm) 2 - 4 Emerson Process Management Manufacturing GmbH & Co. OHG...
Page 26
August 2005 2-3-2 Dual Enclosure Analyzer Standard-Ausführung Standard Version Ausführung „erhöhter Innendruck“ Version „increased pressure“ with category 2 purging system Fig. 2-2b Dual enclosure analyzer (approx. dimensions in mm) Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 5...
The purpose(s) for which the differential pressure switch is used (i.e. to disconnect power or to sound an alarm or otherwise maintain the safety of the installation) is the responsibility of the user! 2 - 6 Emerson Process Management Manufacturing GmbH & Co. OHG...
Page 28
Read and take into account the separate flame arrestor documentation, shipped together with the instrument! Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 7...
100 F Gas Flow Diagrams ® 2-4-3 BINOS 100 F Gas Flow Diagrams 2-4-3-1 Gas Flow Diagram for Nonflammable Gases or Flammable Gases diluted below 1/4 LEL Fig. 2-4d Gas Flow Diagram Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 11...
Page 33
MLT 2 / BINOS 100 F August 2005 ® 2-4-3 BINOS 100 F Gas Flow Diagrams 2-4-3-2 Gas Flow Diagrams for Nondiluted Flammable Gases Fig. 2-4e Gas Flow Diagram 2 - 12 Emerson Process Management Manufacturing GmbH & Co. OHG...
Page 34
MLT 2 / BINOS 100 F August 2005 ® 2-4-3 BINOS 100 F Gas Flow Diagrams 2-4-3-2 Gas Flow Diagrams for Nondiluted Flammable Gases (cont’d) Fig. 2-4f Gas Flow Diagram Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 13...
Page 35
MLT 2 / BINOS 100 F August 2005 ® 2-4-3 BINOS 100 F Gas Flow Diagrams 2-4-3-2 Gas Flow Diagrams for Nondiluted Flammable Gases (cont’d) Fig. 2-4g Gas Flow Diagram 2 - 14 Emerson Process Management Manufacturing GmbH & Co. OHG...
For instructions on how to change the no explosive atmosphere in codes see the separate pressurization the vicinity of the unit (fire instruction manual! safety certificate). Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 15...
(minimum distance to non-IS cables etc.). If running in minimum distance is not possible clearance may be achieved utilizing so- lid insulation of suitable thickness. 2 - 16 Emerson Process Management Manufacturing GmbH & Co. OHG...
MUST be shut off too! This ensures that no hazardous voltages are present within the analyzer enclosure when not pressurized. Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 17...
16, 18 and PE (fig. 2-4). meets the analyzer nominal voltage! Terminal 16 Terminal 18 PE terminal Fig. 2-5 Terminal location inside control unit 2 - 18 Emerson Process Management Manufacturing GmbH & Co. OHG...
This alarm loop operates in normally closed operating conditions. connection mode. Consult factory for applicable circuitry which is available from various manufacturers. Fig. 2-7 Pressure monitor: Terminal allocationand signal connection 2 - 20 Emerson Process Management Manufacturing GmbH & Co. OHG...
Cable gland hexagon socket screw sealing plug Switching On In a next step the analyzer may be powered by opening the purge gas supply and switching on the power supply. Emerson Process Management Manufacturing GmbH & Co. OHG 3 - 1...
Page 43
ATEX Instruction Manual ETC01035 ® MLT 2 / BINOS 100 F January 2004 3 - 2 Emerson Process Management Manufacturing GmbH & Co. OHG...
Refer to the associated manuals for detailed the safe operation of your information maintenance process at risk and affect the replacement of parts. explosion protection! Emerson Process Management Manufacturing GmbH & Co. OHG 4 - 1...
Connect an external source of lines. • compressed air as described in the drawing on the next page (fig. 4-1). 4 - 2 Emerson Process Management Manufacturing GmbH & Co. OHG...
(sec. 4-4-1-3). the measured value is acceptable compared to the following values. Acceptable flow values are: • single analyzer: 3 l/minute maximum • dual analyzer: 4.5 l/minute maximum Emerson Process Management Manufacturing GmbH & Co. OHG 4 - 3...
Connect an external source of lines. • compressed air as described in the drawing on the next page (fig. 4-2). 4 - 4 Emerson Process Management Manufacturing GmbH & Co. OHG...
(sec. 4-4-2-3). the measured value is acceptable compared to the following values. Acceptable flow values are: • single analyzer: 3 l/minute maximum • dual analyzer: 4.5 l/minute maximum Emerson Process Management Manufacturing GmbH & Co. OHG 4 - 5...
August 2005 4-4 Verifications and Tests 4-4-3 Removal of Modifications Remove all the modifications described in section 4-4-1 resp. 4-4-2. Take special care of the gas connections to be tightend. 4 - 6 Emerson Process Management Manufacturing GmbH & Co. OHG...