Data Input
SBI Commands (Data Input Format)
The computer connected via the data port can send control commands to the
balance to control balance and application program functions�
These control commands may have different formats and contain up to 20
characters� Each of these characters must be sent based on the setup configuration
for data transmission�
Formats for Control Commands (Syntax)
Format 1:
Esc
!
CR
Format 2:
Esc
!
#
Esc:
Escape
!:
Command character
#:
Digit
&:
Parameter (number or letter)
_:
Underline (ASCII: 95)
CR:
Carriage return (optional)
LF:
Line feed (optional)
Examples:
Format 1: Esc P
Format 2: Esc x1_
LF
_
CR
LF
Cubis MSE Operating Instructions
Data Interfaces
71