Example: Determining residual weight in percent
Parameter settings: Applic. - percent
Reference percentage value: Ref 100%
Step
 1� Tare the balance
 2� Information:
Enter the reference percentage
(Changing the reference: see the previous page)
 3� Place sample equal to 100% on the balance
(in this example: 111�6 g)
 4� Initialize the balance
The current reference weight remains
saved until a new reference is set
 5� Remove sample (e�g� for drying)
 6� Place unknown weight on balance
(in this example: 105�9 g)
 7� If desired, print the percentage
 8� Toggle display between weight and percentage
 9� Clear display of residual weight and reference percentage
10� If desired, print the residual weight
Press key
J
S
V
P
Repeatedly press S
F
P
Cubis MSE Operating Instructions
Application Programs
Display/Printout
0.0 g
%
ref 100
+ 100.0
 
%
pRef
100 %
Wxx%
+
111.6 g
+ 94.9
%
 
Prc
+
94.9 %
+ 105.9 g 
+ 94.9
%
 
+ 105.9 g
N
+
105.9 g
51