Peressigsäure-Konzentration
Berechnungsformel:
1
c (PES) (in ppm)
= Verbrauch B an Thiosulfatlösung (in ml)
* 38 * N * 1000/(Probevolumen (in ml))
N = Normalität der Thiosulfatlösung
Beispiel:
Verbrauch B = 10,6 ml; c (Thiosulfatlösung) = 0,02 N;
Probevolumen = 20 ml
damit ist
c (PES) = 10,6 x 38 x 0,02 x 1000/20 = 403 ppm oder 0,0403 %.
HINWEIS
1
) ppm = parts per million [Teile pro einer Million Teile]
1 ppm = 1 mg/1 000 000 mg = 1 mg/1000 g = 1 mg/l (1 l H
O = 1 kg)
2
1 % = 1 g/100 ml = 10 g / 1000 ml = 10000 ppm