NEW BRUNSWICK SCIENTIFIC BioFlo 415 Manual To Operations page 55

Benchtop sip fermentor
Table of Contents

Advertisement

6a. Press
pH here...
8. When the
CURRENT VALUE
reading stabilizes,
press the
SET ZERO button.
9. Rinse the pH electrode with distilled water.
10. Immerse pH electrode into a second pH buffer solution which is several pH units
above or below pH 7.00 (e.g., pH 4.00) and allow a few minutes for the system to
equilibrate.
11. Similar to step 7 above, touch the
opens to enter the value of the second buffer solution (e.g., 4.00), then press the
OK
button.
12. When the
13. To ensure accuracy, repeat Steps 4-11 a few times, using the same two buffer
solutions.
NOTE:
The pH calibration should be checked after sterilization, immediately prior
to inoculation. Take a sample from the vessel and compare the pH value
displayed on the control cabinet screen to the pH recorded by an external
pH meter. Any discrepancy should be adjusted with the
procedure.
New Brunswick Scientific
Figure 17: Calibration Screen (pH)
CURRENT VALUE
reading stabilizes, press the
SET SPAN
edit box. Use the touchpad that
6b. ...and pH
appears here.
7. Touch the SET
ZERO edit box. Use
the touchpad that
opens to enter 7.0,
then press the
touchpad's OK
button.
SET SPAN
button.
SET ZERO
User's Guide
43