14
After the vessel and cabinet are in place, read and follow the instructions below, with
reference to Figures 3a & 3b (noting the CAUTION), to install the sparge and
exhaust assemblies and the probes:
VESSEL LIGHT
LEVEL
12mm SPARE PORT
(SINGLE SEPTUM)
12mm PORT
HARVEST/SAMPLE
12mm PORT
HARVEST/SAMPLE
CAUTION!
Never use a wrench on headplate ports; only finger tighten the adapters.
If a leak should be detected at a port, replace the O-ring.
BioFlo 415 M1360-0050
Figure 3a: 7L Headplate Diagram
THERMOWELL
EXHAUST
LEVEL
FRONT of VESSEL
SPARGER
(4) C'SINK HOLES FOR
BAFFLE PLUGS
12mm SPARE PORT
(SINGLE SEPTUM)
19mm SPARE PORT
(MULTIPLE SEPTUM)
OPTIONAL MULTI-
PORT SEPTUM
PG 13.5 PORT
DO/pH
PG 13.5 PORT
DO/pH
12mm SPARE PORT
(SINGLE SEPTUM)
User's Guide