NEW BRUNSWICK SCIENTIFIC BioFlo 415 Manual To Operations page 31

Benchtop sip fermentor
Table of Contents

Advertisement

2. Connect the top of the exhaust condenser line (see Figure 5) to the cabinet
at the location labeled
NOTE:
Ensure that the exhaust filter housing and the tri-clamp gasket are in
place, then make sure that the tri-clamp is secured.
3. Connect the condensate return lines from the inlet filter (see Figure 4) and
exhaust filter (see Figure 5) to the cabinet at the location labeled
CONDENSATE
4. Connect the Control Cabinet's
connect on the vessel jacket (see Figure 7). This is used to supply water to
the vessel jacket.
5. Connect the Control Cabinet's
connect on the vessel jacket (see Figure 7). This is used to channel water
out of the jacket.
WATER OUTLET
To lower exhaust
condenser water inlet
WATER INLET
6. With reference to Figures 7 (above) and 5, connect the
inlet "T" to the exhaust condenser's lower connection.
New Brunswick Scientific
UPPER CONDENSER
.
Figure 7: Vessel Water Inlet/Outlet
.
LOWER JACKET WATER
UPPER JACKET WATER
19
FILTER
to the lower quick-
to the upper quick-
FRONT of VESSEL
's water
JACKET
User's Guide