Gauge Screens - NEW BRUNSWICK SCIENTIFIC BioFlo 415 Manual To Operations

Benchtop sip fermentor
Table of Contents

Advertisement

LEVEL PROBE: color
indicates probe status
Red = DRY
Green = WET
LOOP: each gauge
indicates setpoint
(SP) and process
variable (PV).
Title color indicates
loop status:
Red = OFF
Green = ON
Blue = MANUAL.
Touch loop gauge to
go to full gauge
screen (see Section
6.2.5).
6.2.5
The screen name and
icon in the upper
lefthand corner remind
you that this is a
"Gauge Screen"
LoopName:
Decimal Places:
Press the appropriate
button to display
values with 0, 1, 2 or
3 decimal places.
New Brunswick Scientific
Figure 12: Sample Synoptic Screen

Gauge Screens

Every loop has its own gauge screen. To access it, in the
touch the blue box in the
loop. Your touch will open that loop's
Figure 13: Sample GAUGE Screen
VALVE or HEATER: color indicates valve status:
red = CLOSED, green = OPEN.
Touch valve icon to go to STERILIZATION screen.
LoopName
column that corresponds to the desired
GAUGE
PHASE INDICATOR
indicates the current
phase of operation
PUMPS: each
gauge indicates
setpoint (SP)
and process
variable (PV).
Pump icon and
gauge title color
indicates that
pump's status:
red = OFF,
green = ON
SUMMARY
screen (see Figure 13).
Units:
The action of this
loop, Agitation, is
measured in
RPMs.
Limits:
Here you adjust
the high & low
settings for this
specific loop.
When adjusted,
the scaling for the
gauge will also be
adjusted to reflect
the high & low
limits selected.
User's Guide
31
screen,

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents