Step 5. Verify Operation Of The Network Analyzers; The Operator's Check - Keysight Streamline Series Startup Manual

Vector network analyzers
Hide thumbs Also See for Streamline Series:
Table of Contents

Advertisement

Getting Started

STEP 5. Verify Operation of the Network Analyzers

STEP 5. Verify Operation of the Network Analyzers

The Operator's Check

Allow the P937xA network analyzer to warm up for at least one hour before
performing the Operator's Check.
The operator's check is a software driven test that checks the basic operation
of the assemblies in all of the measurement port signal paths. By performing
the operator's check, the following are determined:
— attenuation ranges of all installed attenuators
— calibration of the receivers
— frequency response of the receivers
— phase lock
— noise floor and trace noise
— various voltages throughout the network analyzer
Accessories Used in the Operator's Check
Male short(s) or open(s) from a 3.5 mm calibration kit.
Performing the Operator's Check
Perform the Operator's Check on each P937xA network analyzer in your
configuration.
1. On your PC screen, click the Network Analyzer icon.
2. Select a P937xA network analyzer in your configuration.
3. On the soft front panel, select Utility > System > Service > Verification >
4. In the P937xA Operator's Check dialog box (refer to
5. Select Begin.
6. If shorts and opens are not connected to all ports, you will be prompted to
7. The result of the operator's check will be shown as a PASS or FAIL next to
1-20
Operator's Check.
page
21), under Configure, select either Prompt for attachment of
Short/Open, to pause at each step in the process to allow moving the
short/open to the appropriate port, or Shorts/Opens are attached to ALL
ports, to run through the test without stopping. Shorts and opens can be
mixed on the test ports.
connect them as they are needed.
each test (refer to
Figure 1-12 on page
Figure 1-12 on
21.
Keysight Technologies Startup Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

P9370aP9371aP9372aP9373aP9374aP9375a