Alarm List - Siemens SITRANS SL Operating Instructions Manual

Continuous gas analysis, in situ laser gas analyzers
Hide thumbs Also See for SITRANS SL:
Table of Contents

Advertisement

Alarm, error, and system messages

10.2 Alarm list

10.2
Alarm list
SITRANS SL can produce alarms from the following list.
Symbol/Group
Error
Code
E6
E8
E45
E46
E47
E48
E66
E68
170
Alarm (description)
Startup procedure active
Line locking failure
FPGA timeout receiver
FPGA timeout receiver junction
FPGA timeout transmitter
Laser unpowered
Curve fit algorithm
Line locking failure
Explanation
/Action
The system is starting up. This alarm should disappear
when system has started.
During normal ambient conditions the SITRANS SL
should have started up within 5 minutes. Ensure that the
ambient conditions are within the specifications for the
SITRANS SL and restart the system if needed.
The SITRANS SL has failed to lock on to the built in
reference cell.
Ensure that the ambient conditions are within the
specifications for the SITRANS SL and restart the
system. If the fault remains, contact Siemens support.
Internal communication failure in the Receiver.
Restart the system. If the fault remains, contact Siemens
support.
Internal communication failure in the Receiver.
Restart the system. If the fault remains, contact Siemens
support.
Sensor connection cable communication failure.
Check that sensor connection cable is properly
connected and not damaged.
The SITRANS SL has stopped measuring to protect the
laser from getting too hot.
The system may have been exposed to an ambient
temperature higher than permitted. Check that the
ambient temperature is within the specifications and
ensure that the system is not exposed to direct sunlight.
Restart the system.
The curve fit algorithm fails due to laser drift.
Ensure that the ambient conditions are within the
specifications for the SITRANS SL and restart the
system. If the fault remains, contact Siemens support.
The line locking state machine ended up in a fault state.
Restart the system. If the fault remains, contact Siemens
support.
Operating Instructions, 12/2010, A5E01132948-04