Siemens SITRANS SL Operating Instructions Manual page 52

Continuous gas analysis, in situ laser gas analyzers
Hide thumbs Also See for SITRANS SL:
Table of Contents

Advertisement

Installing
4.3 Electric connections non Ex
Connection cable
Table 4- 3
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
Chassis
Chassis
1)
2)
3)
4)
5)
54
Electric connections for analog I/O and Modbus connection cable, valid for non Ex
product versions
Terminals in receiver junction board
+
-
Closed when energized
Closed when energized
+
-
+
-
+
-
Modbus D1 (RxD/TxD_N - data
inverted))
Modbus D0 (RxD/TxD_P - data not
inverted)
Modbus Shield
Tx+
Tx-
Rx+
Rx-
+
-
+
-
Maximum possible power consumption.
Maximum input values.
Maximum output values. The following figures show how the outputs are used.
Supported relay modes: "Normally Energized" (default), "Normally De-Energized". For
configuration, see Configuration (Page 116).
We recommend not to connect the connection cable directly to Ethernet connectors.
Instead the sensor connection cable kit should be used for this purpose.
Function/Power
Supply voltage
19 ... 30.2 V DC, 10 VA
Digital OUT 0 (relay)
4)
30 V, 0.5 A
Digital OUT 1 (relay)
4)
30 V, 0.5 A
Digital IN 0
0 ... 30 V DC
Analog OUT 0 (Measurement)
30 V DC, 24 mA
Analog OUT 1 (Measurement)
30 V DC, 24 mA
RS-485 Modbus
- 7 ...+ 12 V DC
Ethernet
5)
Analog IN 0 (Temperature)
0 .. 30mA
2)
Analog IN 1 (Pressure)
0 ... 30 mA
Ground
Ground
Ground
Ground
Operating Instructions, 12/2010, A5E01132948-04
Ethernet cable
1)
3)
3)
2)
3)
3)
White/Orange
Orange
White/Green
Green
2)
Shield
SITRANS SL