Siemens SITRANS SL Operating Instructions Manual page 73

Continuous gas analysis, in situ laser gas analyzers
Hide thumbs Also See for SITRANS SL:
Table of Contents

Advertisement

Connection cable
Table 5- 4
Electric connections for connection cable analog and Modbus, valid for ATEX product versions
Connector position
Conductor color
(Receiver junction board)
Red
1
Blue
2
Pink
3
Grey
4
White
5
Brown
6
Green
7
Yellow
8
Black
9
Violet
10
Grey - pink
11
Red - blue
12
White - green
13
Brown - green
14
White - yellow
15
White - grey
20
Grey - brown
21
White - pink
22
Pink - brown
23
Yellow - brown
Chassis
Shield
Gland
1)
2)
3)
4)
SITRANS SL
Operating Instructions, 12/2010, A5E01132948-04
Maximum possible power consumption.
Maximum input values.
Maximum output values. The following figures show how the outputs are used.
Supported relay modes: "Normally Energized" (default), "Normally De-Energized".
For configuration, see section Configuration (Page 116).
Remarks
Function
+
Supply voltage
19 ... 30,2 V DC, 10 VA
-
Closed when
Digital Out 0 (relay)
energized
30 V DC, 0.5 A
4)
Closed when
Digital Out 1 (relay)
energized
30 V DC, 0.5 A
4)
+
Digital In 0
0 ... 30 V DC
-
+
Analog Out 0
30 V DC, 24 mA
-
+
Analog Out 1
30 V DC, 24 mA
-
Modbus D1
(RxD/TxD_N)
RS 485 (Modbus)
Modbus D0
-7 ... + 12 V DC
(RxD/TxD_P)
Shield Modbus
+
Analog In 0
0 ... 30 mA
-
+
Analog In 1
0 ... 30 mA
-
PE
PE
Explosion protection
5.1 ATEX
1)
3)
3)
2)
3)
3)
2)
2)
75