Connecting The Connecting Cable; Supply Voltage; Cable Entry - Endress+Hauser Proline Prosonic Flow 91W Technical Information

Ultrasonic flow measuring system. flowrate measurement for standard applications with drinking water and process water.
Hide thumbs Also See for Proline Prosonic Flow 91W:
Table of Contents

Advertisement

Connecting the connecting
cable

Supply voltage

Cable entry

8
Connection sensor cable in the electronics compartment
a
b
a, b
Sensor connection cable
c
Cable gland holder
d
Cable gland
Transmitter
85 to 260 V AC, 45 to 65 Hz
20 to 55 V AC, 45 to 65 Hz
16 to 62 V DC
Sensor
Powered by the transmitter
Power supply and signal cables (inputs/outputs)
• Cable entry M20 × 1.5 (8 to 12 mm; 0.31 to 0.47 in)
• Cable gland for cables, 6 to 12 mm ( 0.24" to 0.47")
• Thread for cable entries ½" NPT, G ½"
Connection cable (sensor /transmitter)
Cable glands for one multicore connection cable (1 ×
• Cable gland M20 × 1.5
• Thread for cable entries ½" NPT, G ½"
Connection cable (sensor /transmitter)
Cable glands for two single core connection cable (2 ×
• Cable gland M20 × 1.5
• Thread for cable entries ½" NPT, G ½"
The Prosonic Flow W DN 15 to 65 (½ to 2½") is grounded via the cable gland.
Cable gland for one multicore connection cable (1 ×
d
8 mm) per cable entry
4 mm) per cable entry
8 mm /0.31 in) per cable entry
Proline Prosonic Flow 91W
c
A0015907
a0016008
Endress+Hauser