ProMinent multiFLEX M10 User Manual page 54

Table of Contents

Advertisement

Electrical
AC Input
Fusing
Surge-Spike Suppression
AC Terminals
Sensor, Digital Input
Terminals
DC Loop – Turbine Meter
Power
Mechanical
Enclosure
Rating, Dimensions,
Weight
120VAC Plug Boxes
Environmental
M714_User_Master.doc
Updated Aug. 25, 2015
Rating - Detail
120 or 240 VAC, 50/60Hz,
M10 Version
12.6 Amps @ 120VAC
6 Amps @ 240VAC
M5 Version
6.3 Amps @ 120VAC
3 Amps @ 240VAC
Relays 2-5 and 7-10, NO contacts
snubbed 0.1uF, 150R
Varistor on control AC input
AC Input: AWG 12, 240mm
AC Outputs: AWG 14, 150mm
AWG 22, 0.25 – 0.50mm
15 – 22 VDC, unregulated
Thermally fused @ 200mA
Rating
Non-metallic, NEMA4X, IP65
M10 Version
14"W x 17"H x 6.75"D
355mm W x 430mm H x 170mm D
17lb. 7.7kg nominal
M5 Version
12.5"W x 15"H x 6.5"D
320mm W x 380mm H x 165mm D
11.8lb. 5.4kg nominal
Rated for outdoor use.
Limited to 5 Amps / plugbox
2 & 4 plug boxes provided with 36"
of flexible non-metallic conduit
Pollution Degree 2,
Altitude 2000m,
Installation Category II,
Humidity 5% to 95%,
S p e c i f i c a t i o n s
4 . 3
Switch selectable
5x20mm, 120VAC fusing:
Relays 1-5 & 6-10: 6.3A ea.
Control: 1.5A
Controller, transformer isolated
from AC line
2
Electrical grounds at bottom of
aluminum backplate
2
Conductor insulation rated
600VAC minimum.
2
MAX AWG14, 150 150mm
Conductor insulation rated
600VAC minimum.
Field wiring terminals
on M7 & M14
Nominal dimensions, excluding
entry fittings and flexible
conduit. Including mounting.
Enclosure door hinged left.
Allow 18", left for door opening.
Allow 24", below for cable-
conduit access.
Plug boxes not included:
1. At hardwired and
2. With controllers shipped
4 o f 4
Notes
2
Notes
240VAC sites.
outside of North America
54

Advertisement

Table of Contents
loading

This manual is also suitable for:

Multiflex m5

Table of Contents