Reinstall the Echo Liquid Handler after a move

After the move, use the following procedure to restore an Echo Liquid Handler to normal operation.
Note: The steps below apply to a standalone configuration of the Echo Liquid Handler. If the Echo Liquid Handler is
integrated into an Access system, please contact Labcyte Service and Support. For more information, see
Information.
1. Ensure that the new mounting surface is level.
2. Set up the Echo Liquid Handler, bottles, computer, and other ancillaries. Remember to allow for minimum operating space
around the Echo Liquid Handler. For more information, see
3. Set up and start the chiller. For more information on Chiller setup, see
4. Set up the coupling fluid bottle. Ensure that the bottle cap is firmly attached to ensure a tight vacuum seal. Place the filled
coupling fluid bottle in the caddy. The fluid level sensor (red LED) should be visible on the side wall of the instrument to
confirm that the bottle is positioned correctly and is filled.
Figure 38: Coupling fluid level sensor in bottle caddy
Note: Coupling fluid bottle should be pressed against the sensor located in the adjustable metal bracket to
ensure that it works correctly.
5. Connect the air pressure and vacuum pressure supply lines to the fluidics panel.
Figure 39: Bottle connections
Caution: Do not run the pump until the vacuum source is connected and active. Running the pump without
vacuum can lead to flooding the system or a system error.
75
Site
Requirements.
Chiller
Setup.
Instrument Relocation
Contact
PN | 001-11665