Instrument Specifications; Communications; Fluid Transfer - Labcyte Echo 525 User Manual

Rapid, acoustic liquid handler
Table of Contents

Advertisement

LABCYTE INC.
CHAPTER 7 | Instrument Specifications

7 | Instrument Specifications

Each Echo Liquid Handler conforms to the specifications listed below.

Communications

l

Fluid Transfer

l
Labware
l
Electrical
l
Ratings
l
Physical
l
Environmental
l
Client PC (minimum requirements)
l
Optional Hardware
l
Communications
Server connection: Echo Liquid Handling Platform (Server Version 2.6.x)
Network connection: 10/100/1000 BaseT
Network protocol: TCP/IP
User account: Labcyte (The client PC must have a Labcyte account, with administrator privileges, that is accessible by Labcyte
Service and Support to service the Echo Liquid Handler system.)
Fluid Transfer
Drop transfer resolution: 25 nL
Volume transfer range:
25 nL to 5 µL (one well to one well)
25 nL to 45 µL (one well to many wells)
25 nL to Working Range (one to many wells)
Note: Actual transfer range is dependent on fluid type and assay plate.
Table 15: Fluid Transfer Ranges by Plate Table
Plate
384PP
384PP Plus
384LDV Plus
6RES
Transfer Accuracy: <10% deviation from target volume
Transfer Precision: <8% CV
USER GUIDE|Echo® 525 Liquid Handler User Guide
Maximum
65 µL
65 µL
14 µL
2800 µL
Minimum
15 µL
20 µL
4.5 µL
250 µL
Working Range
50 µL
45 µL
9.5 µL
2550 µL