The consumables and corresponding specifications recommended to be used with the Echo Liquid Handler are listed below:
Table 5: Compatible Source Plates for Echo 525 Liquid Handler
Number of
Wells
Surface
Treated
Volume Range
( µL)
Working
Volume ( µL)
Calibration
384PP_
AQ_BP
384PP_
AQ_SP
Calibration and
Example Fluids
384PP_
AQ_
SPHigh
53
384PP Plus Plate and
384PP Plate
384
No
20-65
45
Example
Calibration
Fluids
buffer with
384PP_
protein
Plus_AQ_
BP
buffer with
384PP_
surfactant and
Plus_AQ_
protein
SP
384PP_
Plus_AQ_
GP
buffer with
384PP_
surfactant and
Plus_AQ_
protein,
GPSA
surfactant >=
50% CMC)
CMC (Critical
Micelle
Concentration
384PP_
Plus_AQ_
GPSB
384PP TC Plate
384
Yes
20-65
45
Example
Calibration
Fluids
buffer with
protein
buffer with
surfactant
and protein
buffer with
glycerol
384LDV_
and protein
AQ_GP
buffer with
glycerol,
protein,
and
surfactant;
surfactant
5-14% CMC
buffer with
glycerol,
protein,
and
surfactant;
surfactant
14-50%
CMC
Component Description
384LDV Plus Plate
384
Yes
4.5-14
9.5
Example
Calibration
Fluids
6RES_AQ_
BP2
6RES_AQ_
GPSA2
buffer
with
glycerol,
protein
6RES__
AQ_GPSB2
Echo Qualified
Reservoir
6
Yes
250-2,800
2,550
Example
Fluids
buffer 0 -
30% V/V
glycerol
and 0-
2.5%
CMC
surfactant
buffer 0 -
30% V/V
glycerol
and 2.5-
30% CMC
surfactant
buffer
with 0 -
30% V/V
glycerol
and 30-
50% CMC
surfactant