Component Description

AC Power and Fuse Compartment

The AC power and fuse compartment (as called the power input module) is located on the rear panel, to the left of the Emergency
Motion Off button. It contains the main Echo Liquid Handler power on/off switch, AC power outlet, and fuse compartment.
The Echo Liquid Handler instruments are built in two AC power configurations:
Table 3: AC Power Configurations for Echo Liquid Handlers
Configuration
A
100-120 V
B
200–240 V
The fuse compartment for the Echo Liquid Handler contains two AC power fuses. If a fuse requires replacement, contact Labcyte
Service and Support (see
Contact
Figure 24: AC power and fuse compartment

Ethernet Connection

The Echo Liquid Handler includes several communication connections to the right of the Emergency motion off button. The RJ-45
Ethernet port is the only data connection required for normal Echo Liquid Handler operation. The Ethernet cable connects the
Echo Liquid Handler directly to the client PC or to a network. This connection allows you to monitor and control the Echo Liquid
Handler. If the Echo Liquid Handler is connected directly to a client PC, the Ethernet cable must be a "cross-over" cable.
Other data connection ports are used by Labcyte authorized personnel.
Figure 25: External connections: power and data
Voltage
Frequency
50/60 Hz
50/60 Hz
Information).
CHAPTER 3 | General Description
Power
10 A
115 VAC, 10 A, FAST (P/N 001-3966)
5 A
250 VAC, 5 A, FAST (P/N 001-3967)
Fuse
38