USER GUIDE | Echo® 525 Liquid Handler User Guide
Maintenance Schedule
Echo Maintenance Alerts
The Echo Liquid Handler contains internal checks to ensure that it performs liquid fills under optimal conditions. If, for example, the
coupling fluid temperature is outside the normal range (21.2°C – 23.0°C), a yellow warning light appears on the Echo front and back
panels. If the coupling fluid temperature continues to rise or fall outside of the warning range (20.2°C – 24.0°C), a red fault light
appears.
When you see the warning or fault lights, click the Diagnostics tab in the Echo Liquid Handler software to view the specific alert.
Figure 102: Diagnostic window
149
PN | 001-11665