Echo Maintenance Alerts - Labcyte Echo 525 User Manual

Rapid, acoustic liquid handler
Table of Contents

Advertisement

USER GUIDE | Echo® 525 Liquid Handler User Guide
Maintenance Schedule

Echo Maintenance Alerts

The Echo Liquid Handler contains internal checks to ensure that it performs liquid fills under optimal conditions. If, for example, the
coupling fluid temperature is outside the normal range (21.2°C – 23.0°C), a yellow warning light appears on the Echo front and back
panels. If the coupling fluid temperature continues to rise or fall outside of the warning range (20.2°C – 24.0°C), a red fault light
appears.
When you see the warning or fault lights, click the Diagnostics tab in the Echo Liquid Handler software to view the specific alert.
Figure 102: Diagnostic window
149
PN | 001-11665

Hide quick links:

Advertisement

Table of Contents
loading