Labware; Electrical; Ratings; Physical - Labcyte Echo 525 User Manual

Rapid, acoustic liquid handler
Table of Contents

Advertisement

USER GUIDE | Echo® 525 Liquid Handler User Guide

Labware

Source Labware Compatibility: 6RES (ER-0050), 384PP (P-05525), 384PP TC (PPT-05525), 384PP Plus (PL-05525) , 384LDV Plus
(LPL-0200)
Destination Labware Compatibility: All Echo qualified microplates and most ANSI-compliant/SBS-standard microplates in 96-well
and 384-well format that are 8-16 mm in height are compatible. Use of 1536-well plates as destination plates is application
dependent.
The table below shows the approximate fill time per plate transferred per 24 hour period, including robotic movement of
microplates in and out of the instrument.
Table 16: System Throughput
Source Plate Type
Reservoir (6RES)
Reservoir (6RES)
384PP/384PP Plus
384PP/384PP Plus
384PP/384PP Plus
384PP/384PP Plus
384LDV Plus
384PP Plus
384PP Plus

Electrical

There are four electrical configurations of the Echo Liquid Handler:
Table 17: Electrical Configurations for the Echo Liquid Handler
Voltage
100-120 V
50 Hz
100-120 V
60 Hz
200–240 V
50 Hz
200–240 V
60 Hz
Note: The electrical requirements listed above do not account for chiller power requirements.

Ratings

Noise Rating: 64dBA
l
Power Rating: 600W without the chiller
l
Heat Rating: 600W without the chiller
l

Physical

Weight: 128 kg (283 lbs)
l
Dimensions:
l
165
Destination Plate Type
384
1536
384
384
384
384
384
1536
1536
Frequency
10 A
10 A
5 A
5 A
Dispense Type
1:384
1:1536
384:384
384:384
16:384
48:384
384:384
4x384:1536
4x384:1536
Power
115 VAC, 10 A, FAST (P/N 001-3966)
115 VAC, 10 A, FAST (P/N 001-3966)
250 VAC, 5 A, FAST (P/N 001-3967)
250 VAC, 5 A, FAST (P/N 001-3967)
Dispense Volume Per
Destination Well
5 µL
1 µL
25 nL
1 µL
1 µL
5 µL
100 nL
25 nL
1 µL
Fuse
Labware
Destination Plate Fill
Cycle Time (min)
9.3
17.2
2.2
3.3
3.7
10
3.8
7.9
12.4
PN | 001-11665