Zeiss Humphrey Matrix 800 User Manual page 155

Table of Contents

Advertisement

576
Printer Error! Unable to print System
Information!
584
Unable to retrieve login credentials
for __________________
593
Failure to export exam to DICOM
archive
619.2
Critical Error! Factory calibration
needed. Select 'Shut Down' and then
cycle power to restart your
instrument and restore normal
operation. If problem persists,
contact Customer Service. See Help
F6.
619.3
Critical Error! Factory calibration
needed. Select 'Shut Down' and then
cycle power to restart your
instrument and restore normal
operation. If problem persists,
contact Customer Service. See Help
F6.
619.4
Critical Error! Factory calibration
needed. Select 'Shut Down' and then
cycle power to restart your
instrument and restore normal
operation. If problem persists,
contact Customer Service. See Help
F6.
702
Unable to connect to DICOM server.
704
One or more patients could not be
imported.
Humphrey Matrix User Manual
Numerous printer errors:
• Not connected
• No paper in tray
• Non-approved printer
• No ink
User attempted to perform the
pattern test 'Set Pattern' from the
Advanced Diagnostics screen. This
operation could not complete
because their system needs to be
factory calibrated.
User attempted to perform an
'Ambient Light Level' check from the
Advanced Diagnostics screen. This
operation could not complete
because their system needs to be
factory calibrated.
User attempted to perform a
'Monitor Brightness' check from the
Advanced Diagnostics screen. This
operation could not complete
because their system needs to be
factory calibrated.
Some of the patients could not be
imported from the serial EMR
application.
Error Codes
Verify the printer attached to the
system is an approved printer.
Perform the print sample test from
the Diagnostics screen.
If problem persists, cycle the printer
power or Shut Down and reboot the
Matrix instrument.
If problem still persists, try a different
approved printer.
Try again. Check network.
Check DICOM archive. Check DICOM
configuration in System Settings.
SW detected a problem that requires
Shut Down and restart to clear.
Service will be needed if this issue is
chronic and prevents normal
operation of the system.
SW detected a problem that requires
Shut Down and restart to clear.
Service will be needed if this issue is
chronic and prevents normal
operation of the system.
SW detected a problem that requires
Shut Down and restart to clear.
Service will be needed if this issue is
chronic and prevents normal
operation of the system.
Check DICOM archive. Check DICOM
configuration in System Settings.
Check the EMR connection and
setup.
2660021138935 Rev. F 2015-11
D-3