Zeiss Humphrey Matrix 800 User Manual page 153

Table of Contents

Advertisement

Code
Message
Number
(SBC V06.00.0X)
115
Not enough data for Overview found.
Select a patient with enough data.
123
Please limit input to
CHARACTER_LIMIT characters.
148
The default folder
DEFAULT_FOLDER_NAME contents
will be merged but the default folder
cannot be deleted.
151
Invalid folder name entered. Please
enter a valid folder name.
160
Invalid file name entered. Please
enter a valid file name.
167
WARNING! Are you sure you want to
permanently delete the selected test
from the database?
168
WARNING! Selected test will be
permanently deleted from the
database. Continue?
175
Patient not found in database. Add
New Patient?
176
Unable to run test during calibration.
Wait until calibration is completed.
192
Sample test is not present on the
system. Function unavailable.
196
What would you like to do with these
tests?
(D) Error Codes
Error codes that require additional user information have been included in this table. Those
that are self-explanatory have not.
A minimum of 2 visual fields of the
same test type are required for an
overview plot.
Merging any folder with the Main
(default) folder causes this message.
Continue?
Empty folder name field.
Empty file name field when saving
Overview analysis plot as JPEG.
User confirmation needed.
User confirmation needed.
User confirmation needed.
All testing operations are disabled
during calibration.
• The file
/root/GUI/printSample.fdt2
is missing
• The file
/root/GUI/printSample.fdt2
could not be loaded for various
file I/O reasons.
If the user highlighted the tests and
selects the Edit test option on the F3
screen
Humphrey Matrix User Manual
Possible
Cause
A minimum of two visual fields of the
same test type are required for an
overview plot
Delete characters until the number of
characters used is less than or equal
to the number specified.
Select Continue.
Enter folder name.
Enter valid file name.
Deleting tests from the database is
permanent. Unrecoverable if not
available from a prior backup.
Deleting tests from the database is
permanent. Unrecoverable if not
available from a prior backup.
The automatic search feature did not
find a patient with the name entered
in the database – either new patient
or a data entry error.
Check Calibration status in Help (F6)
Calibration screen.
Calibration takes approximately 15
minutes to complete.
SW error if message doesn't go away
when calibration is completed.
(restart).
Try again. If problem persists, send
customer update (from web site?)
which includes necessary file. If after
upgrade, problem still is present may
need service but if this is the case,
the customer should be seeing other
issues with the system as well.
The user decides to either move the
selected tests to a different folder OR
move the test association to another
patient.
2660021138935 Rev. F 2015-11
Error Codes
D-1