6.2.4 Technical information on relative humidity and temperature probes using SICRAM module
Temperature
Model
HP472ACR
HP473ACR
HP474ACR
HP475ACR
HP475AC1R
HP477DCR
HP478ACR
Common characteristics
Relative humidity
Sensor
Resolution
Temperature drift @20°C
Response time %RH at constant temperature
Temperature
Resolution
Temperature drift @20°C
Important notes:
1) Do not let hands touch the RH sensor.
2) The sensor base is in alumina so it could easily break
3) Storage of the saturated solutions: The saturated solutions must be stored in a dark environment
at a constant temperature of about 20°C with the container well closed inside a dry room.
HD32.1
Application range
sensor
%RH
Pt100
0...100%RH -20...+80 °C
Pt100
0...100%RH -20...+80 °C
Pt100
0...100%RH -40...+150 °C
Pt100
0...100%RH -40...+150 °C
Pt100
0...100%RH -40...+180 °C
Pt100
0...100%RH -40...+100 °C
Pt100
0...100%RH -40...+150 °C
Temperature
±1.5% (0...85%RH)
±2.5% (85...100%RH)
(2 + 1.5% measure)%
@ T= remaining range
Capacitive
0.1%RH
0.02%RH/°C
10sec (10 80%RH; air speed=2m/s)
0.1°C
0.003%/°C
- 70 -
Accuracy
%RH
@ T=15...35 °C
Temp.
±0.3 °C
±0.3 °C
±0.3 °C
±0.3 °C
±0.3 °C
±0.3 °C
±0.3 °C
V2.1