The Operating Program B: Discomfort Analysis; He Perating Rogram B: Discomfort Nalysis; The Unit Of Measurement "Unit - DeltaOHM HD32.1 Operating Manual

Thermal microclimate
Table of Contents

Advertisement

4.2 T
O
HE
PERATING
Connect the probes. Turn on the instrument: After few seconds, the measurement display mode will
appear:
The battery charge symbol and current time are up on the left (for further details, see the par. 10).
The following quantities are reported:
Th:
Air temperature detected at head height (1.7 m for a standing person; 1.1 m for a seated
person)
Tb:
Air temperature detected at abdomen height (1.1 m for a standing person; 0.6 m for a seated
person)
Tn:
Temperature of the net radiometer
Tk:
Air temperature detected at ankle height (0.1 m)
Tf:
Temperature at floor level
P:
Net radiation
DT:
Radiant asymmetry temperature
By pressing SHIFT FNC , a drop-down menu is shown with the following information:
• unit : allows selection of the unit of measurement
data : allows display of the maximum, minimum, and average values

4.2.1 The unit of measurement "Unit"

Proceed as follows to access unit function:
Use SHIFT FNC to open the drop-down menu;
• Use the arrow keys ▲▼ to select unit;
• Press ENTER to confirm: the selected quantity is displayed in the central line of the
display;
• The three different temperature units of measurement are displayed in the bottom line of the
display, using F1 , F2 or F3 : The unit is selected and displayed near the value shown in the
central line;
• Press ESC to exit the drop-down menu without making any change.