Logging Menu; Log Interval - DeltaOHM HD32.1 Operating Manual

Thermal microclimate
Table of Contents

Advertisement

5.2 L
M
OGGING
ENU
Once you enter the main menu by pressing SETUP , to access the Logging menu proceed as
follows:
1. Use the arrow keys ▲▼ to select Logging ;
2. Press ENTER :
The parameter setting submenu for the logging sessions (measured data capture) will be
displayed.
You can set the data capture frequency (Log interval) and the automatic l ogging start (Start/stop
time). The capture interval is the same for all probes.

5.2.1 Log Interval

Use this item to set the LOG interval (interval between two subsequent sample captures): To enter
this setting, proceed as follows:
Once you have accessed the LOGGING submenu (previous par.) use the arrow keys ▲ ▼ to select
Log Interval:
1. Use the arrow keys ▲ ▼ to select the interval duration from 15 seconds to one hour;
2. Press ENTER to confirm and return to the Logging menu;
3. Press ESC to return to the Logging menu without making any change;
4. Press ESC again to return to the main menu;
5. Press SETUP to exit immediately from the menu.
These are the available values: 15 seconds - 30 seconds - 1 minute - 2 minutes - 5 minutes - 10
minutes - 15 minutes - 20 minutes - 30 minutes - 1 hour
Storage interval
15 seconds
30 seconds
1 minute
2 minutes
5 minutes
HD32.1
Storage Capacity
About 11 days and 17 hours
About 23 days and 11 hours
About 46 days and 22 hours
About 93 days and 21 hours
About 234 days and 17hours
Storage interval
10 minutes
15 minutes
20 minutes
30 minutes
1 hour
- 36 -
Storage Capacity
About 1 year and 104 days
About 1 year and 339 days
About 2 years and 208 days
About 3 years and 313 days
About 7 years and 261 days
V2.1