5.2.3 Start/stop time – Automatic start
The logging start and end can be programmed by entering the date and time. When called, the
function suggests the current time plus 5 minutes as the start time: Press <ENTER> to confirm or
set the date and time using the arrow keys. Then you are asked to set the data to end the recording:
By default the instrument suggests the start time plus 10 minutes. The default suggested values are
such to allow the user to setup an instrument ready for measurement.
NOTE: By default the set time is 5 minutes after the current time.
To enter this setting, proceed as follows.
Once you have accessed the LOGGING submenu (previous par.) use the arrow keys ▲ ▼ to select
Start/Stop time: The following message "Enter start time" will be displayed:
1. Use the arrow keys ◄ ► to select the data to be changed (year/month/day and
hour:minutes:seconds);
2. Once selected, the data will start blinking;
3. Use the arrow keys ▼▲ to change its value;
4. Confirm by pressing ENTER ;
6. Press ESC to return to the Logging menu without making any change;
7. Press ESC again to return to the main menu;
8. Press SETUP to exit immediately from the menu.
After setting the logging start time, the logging end time (enter stop time) window will be
displayed:
1. Use the arrow keys ◄ ► to select the data to be changed (year/month/day and
hour:minutes:seconds);
2. Once selected, the data will start blinking;
3. Use the arrow keys ▼▲ to change its value;
4. Confirm by pressing ENTER ;
5. Press ESC to return to the Logging menu without making any change;
6. Press ESC again to return to the main menu;
7. Press SETUP to exit immediately from the menu.