Input Signal
ECG Sensitivity
Range
Respiration Sensitivity
Alarms
Apnea
Low Breath Rate
Bradycardia
Tachycardia
Loose Connection
Low Battery Warning
Low Battery Shutdown
Full Memory
SmartMonitor 2
Non-Volatile Memory
Memory System
Capacity
Logs
Waveform events*
Signal Channels
Internal
Oximeter
External Input
Event Parameters
Internal
Oximeter
Event Duration
Apnea Record
Bradycardia Record
Electrical
Safety Standards
Certification
Power Input:
SmartMonitor 2
Power Input Power
Cord/Battery Charger
Batteries
Physical
Size
Weight
Shipping Weight
10 to 275 BPM, 0.5 mV; 10 to 150 BPM, 0.2 mV
4 to 150 BrPM @ 2 Ohms; 8 to 75 BrPM @ 0.5 Ohm; 30 BrPM
@ 0.15 Ohm
Delay settings of 10, 15, 20, 25, 30, 40 seconds
Off, 4,5, 6, 7, 8, 9, 10, 12,14, 16, 18, 20, 25, 30 BrPM
40, 50, 60, 70, 80, 90, 100 BPM. Optional delay of 5 seconds
Off, 90, 100, 110, 130, 150, 170, 190, 200, 210, 220, 230, 240,
250, 270 BPM
Base Impedance exceeds 2000 ± 200 Ohms
Battery voltage low, power cord/battery charger required
promptly
Battery voltage too low to continue with out Power cord/battery
charger
Memory is 50% full or 80% full, repeats when memory is 100%
full
2 MB
1000 Patient event listings, 1000 Equipment event listings
500 HR trend and impedance; or 100 ECG QRS, HR trend, and
impedance; or 21 hours of continuous HR trend and impedance.
* Oximetry or other signals can be recorded with reduced
capacity
QRS, heart rate trend, and impedance
Saturation and pulse, or pulse rate
Airflow, pH, strain gauge, or other low frequency physiological
signals, +/- 1.25 V maximum
Apnea Record, Apnea Alarm, Low Breath Rate, Bradycardia for
Record, Bradycardia Alarm, Tachycardia Alarm
+3.5V to +5.0V during alarm condition
Alarm duration plus current pre- and post-settings totaling 45, 60,
75, 90 seconds
Off or 6 to 40 seconds in two-second intervals
Off or 50 to 100 Beats per Minute in five-BPM intervals
Meets CSA, C22.2 No. 601-1 and UL2601-1 and EN60601-1
CSA/NRTL
10.5 VDC
120 V, 60Hz or 230 V, 50 Hz
Rechargeable NIMH, 18 – 21 hours continuous use; recharge
time, 6 hours @ 77° F; Shelf life, 30 days if fully charged,
expected battery life is 2 years.
HWD: 4' x 91/4" x 9" (11.2 x 23/5 x 22.9 cm)
2.6 lbs. (1.18 kg)
8.5 lbs. (3.9 Kg)