After the new parameters have been entered into the SmartMonitor 2, the monitor must be turned off to
exit Menu Mode.
To set the parameters through the Synergy-E computer software, refer to the Synergy-E Manual for
instructions on communicating with the SmartMonitor 2.
System Set-up Menu
SmartMonitor 2 also has a System Set-up Menu, which displays the entire menu of parameters. This
includes very seldom-used menu choices. In addition to some menus previously seen, you will have
access to the following menus. To access this menu, do the following:
Press the POWER button ON. After a short delay, the display will read:
INITIALIZING PLEASE WAIT
Then,
MENU MODE? ENTER PROPER KEY SEQUENCE
Press the keys in the following sequence within 10 seconds:
Press the t Up arrow 3 times.
Press the ENTER button 3 times.
Scroll to MENU OPTIONS and press ENTER.
Scroll to SYSTEM MENU and press ENTER.
The following table lists the additional options available from the System menu:
Menu Option
Date Format
Memory Full Alert
Memory Full Audible
Modem Speed -
Monitor
LCD Display
DATE FORMAT
MEMORY FULL ALERT
MEMORY FULL AUDIBLE
Modem Speed-Monitor
Setting Alarm and Recording Limits
Description of Option or Instructions
Permits the selection of two formats for
displaying the DATE:
MONTH/DAY/YEAR or
DAY/MONTH/YEAR
Standard value: Month/Day/Year
Two options are available: 80% full or
50% full.
When the memory usage reaches the
selected limit (50% or 80%), the unit will
generate an alarm.
Standard value: 50%
Used to enable/disable the audible alarm
when the unit reaches the selected
memory almost full limit (50%/80%).
NOTE: This parameter does not affect the
visual indicator; it cannot be disabled.
Values:
ON, OFF
Standard Value: OFF
The default setting is 38,400Bps. This
setting should not be change unless
directed to do so by Respironics
Technical Support.
27