Menu Option
LCD Display
STD Alarm
Are Selected
Parameters
Or
Are Not
Selected
STD Record
Are Selected
Parameters
Or
Are Not
Selected
STD System
Are Selected
Parameters
Or
Are Not
Selected
Apnea Alarm
APNEA ALARM
in seconds
Low Breath
LOW BREATH
Rate Alarm
ALARM BrPM
Bradycardia
BRADYCARDIA
Alarm
ALARM BPM
Bradycardia
BRADY ALARM
Alarm Delay
DELAY in
seconds
Tachycardia
TACHYCARDIA
Alarm
ALARM BPM
Description of Option or Instructions
Press Enter to activate.
Press t or u to change the selection.
When you have your selection displayed
press Enter.
Press Enter to activate.
Press t or u to change the selection.
When you have your selection displayed
press Enter.
Press Enter to activate.
Press t or u to change the selection.
When you have your selection displayed
press Enter.
Establishes the amount of time of no respiration
detection prior to activation of the apnea alarm.
Values:
10, 15, 20, 25, 30, 40 seconds
Standard value: 20 seconds
Establishes the alarm set point based on
frequency of detected respiratory effort.
Values:
OFF, 4, 5, 6, 7, 8, 9, 10, 12, 14, 16, 18, 20, 25,
30 breaths per minute
Standard value: OFF
Establishes the LOW heart rate alarm set point
based on the average detected ECG signal.
Values:
40, 50, 60, 70, 80, 90, 100 beats per minute
Standard value: 80 bpm
Enables a delay to the audible alarm
Values:
0 or 5 seconds
Standard value: 0 seconds
If this parameter is set to 5 seconds, the
audible alarm will not activate unless the alarm
set point is violated for 5 seconds or more.
This is a physician decision and is based on such
factors as the known condition of the patient,
number of short Bradycardia alarms documented
by the SmartMonitor 2, the current Bradycardia
alarm set point, and the patient's average resting
heart rate.
Establishes the HIGH heart rate alarm set point
based on the average ECG signal.
Values:
OFF, 90, 100, 110, 130, 150, 170, 190, 200, 210,
220, 230, 240, 250, 270 beats per minute
Standard value: 230 bpm
Setting Alarm and Recording Limits
Menu Path
All Menus
Alarm Menu
All Menus
Record Menu
All Menus
System Menu
All Menus
Alarm Menu
All Menus
Alarm Menu
All Menus
Alarm Menu
All Menus
Alarm Menu
All Menus
Alarm Menu
21