Respironics SmartMonitor2 Operator's Manual page 10

Professional
Table of Contents

Advertisement

Diagnosis Summary
Diagnosis
Prematurity
RDS, Resp. Failure, HMD
ALTE, Apnea, AOI, AOP
Bronchiolitis, Pneumonia, RSV
Gastro-Esophageal Reflux
Broncho-Pulmonary Dysplasia
Totals by Site
Results for Both Monitors by Site
Apnea
Study Site
with Alarm
Site #1
25
Site #2
2
Site #3
3
Site #4
10
Site #5
21
Site #6
7
68
Totals
Analysis of Results
SmartMonitor
Apnea with Alarm
68
The sensitivity for SmartMonitor is 100*68/142 = 47.89% with 95% confidence limits (39.44%, 56.42%). The
positive predictive value for SmartMonitor is 100*68/165 = 41.21% with exact 95% confidence limits (33.62%,
49.13%).
The sensitivity for SmartMonitor 2 is 100*73/142 = 51.41% with 95% confidence limits (38.25%, 54.30%).
The positive predictive value of Smart Monitor 2 is 100*73/158 = 46.20% with exact 95% confidence limits
(38.25%, 54.30%).
Site # 1
Site #2
10
3
8
0
0
0
Other
5
26
SmartMonitor
No Apnea
with Alarm
without Alarm
21
2
12
21
19
22
97
SmartMonitor 2
Apnea with Alarm
73
Site #3
Site #4
0
0
7
0
0
3
2
8
6
3
0
1
0
0
2
0
0
2
2
5
10
7
13
31
Apnea
Apnea
with Alarm
20
28
3
2
6
3
10
11
25
18
10
11
74
73
Difference
(SmartMonitor 2 – SmartMonitor)
5
Introduction
Site #5
Site #6
13
0
5
0
5
4
0
6
1
1
2
0
10
2
36
13
SmartMonitor 2
No Apnea
Apnea
with Alarm
without Alarm
21
17
2
3
13
6
11
9
19
28
19
6
85
69
% Difference
6.8
Totals by
Diagnosis
30
11
33
10
4
4
34
126
4

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents