Limit Values - ProMinent DULCOMETER D1CA Operating Instructions Manual

Adjustment and operation, measured variable chlorite
Hide thumbs Also See for DULCOMETER D1CA:
Table of Contents

Advertisement

Complete Operating Menu / Description

Limit values

Access to all setting menus can be blocked with an access code !
limits
setting ?
Type of limit transgression
Limit 1:
Limit 2:
Limit value
Limit 1:
Limit 2:
Limit value
Limit 1:
Limit 2:
Switch-on delay ∆t on
Switch-off delay ∆t off
Hysteresis limits
Checkout time limits
18
limit 2
upper
1.50 ppm
limit 1
lower
0.10 ppm
relay 1
LV1
- active closed
∆t on
0 s
∆t off
0 s
Possible values
Initial value
Increment
upper
lower
lower
upper
off *)
0.025 ppm
0.001 ppm
0.375 ppm
0.10 ppm
0.01 ppm
1.50 ppm
0 s
1 s
0 s
1 s
0.010 ppm
0.001 ppm
0.04 ppm
0.01 ppm
off
1 s
limits fault
hyst.:
0.20 ppm
∆t on:
off
control:
on
relay 2
LV2
- active closed
∆t on
0 s
∆t off
0 s
Lower value
Upper value
-0.05 ppm
0.550 ppm
-0.20 ppm
2.20 ppm
0 s
9999 s
0 s
9999 s
0 ppm
0.550 ppm
0 ppm
2.20 ppm
1 s
9999 s
D1C2-Clt-024-GB
Remarks
Limit transgression when
exceeding or dropping below value
*) only with limit relay
for CLT 1-mA-0.5 ppm
for CLT 1-mA-2 ppm
Effective in direction of
"cancelling limit transgression"
Results in message and alarm.
off = 0 s:
Function switched off,
no message, no alarm