Gilson Pipetman Ultra U2 User Manual page 32

Table of Contents

Advertisement

PIPETMAN
Model
(Reference)
Diamond Tip
8x20 (F21040)
Min.
and
12x20 (F21041)
DL10
Max
8x300 (F21042)
Min.
and
12x300 (F21043)
D300
Max. 300
The data given in the tables conform to the ISO 8655-2
Standard.
Each pipette is inspected and validated by qualifi ed
technicians according to the Gilson Quality System.
Gilson declares that its manufactured pipettes comply
with the requirements of the ISO 8655 Standard, by
type testing.
The adjustement is carried out under strictly defi ned
and monitored conditions (ISO 8655-6):
• Basis of adjustment, Ex.
• Reference temperature, 20 °C
• Relative humidity, 50 %
• Barometric pressure, 101 kPa
• Use of distilled water grade 3 (ISO 3696)
• Ten measurements for each test volume, which are
Nominal Volume, 50 % of Nominal Volume, and the
minimum or 10 % of Nominal Volume.
32
Maximum Permissible Errors
Volume
Gilson
(μL)
Systematic
error (μL)
1
± 0.10
2
± 0.10
10
± 0.20
20
± 0.40
20
± 1.00
30
± 1.00
150
± 1.50
± 3.00
u
®
ltra
ISO 8655
Random
Systematic
error (μL)
error (μL)
0.08
± 0.40
0.08
± 0.40
0.10
± 0.40
0.20
± 0.40
0.35
± 8.00
0.35
± 8.00
0.60
± 8.00
1.00
± 8.00
Random
error (μL)
0.20
0.20
0.20
0.20
3.00
3.00
3.00
3.00