the tip ejector button to the most comfortable
position: left, right, or middle.
2) Set the stroke by rotating the adjustment wheel
until you fi nd the position where it is most easy to
activate the tip ejector. You may want to reset the
stroke after fi tting a diff erent type or size of tip.
Tip-ejector Extension for Use with Ultra U2 and U10
In order to eject D10 tips, tip ejector
extensions are supplied with U2 and
U10 pipettes.
The tip-ejector extension which is made
of PVDF (polyvinylidene Fluoride), is
autoclavable.
To fi t a tip-ejector extension:
- hold the pipette with the LCD window uppermost,
- hold the extension with the slot
uppermost,
- slide the extension over the tip-
holder,
- push the extension fi rmly onto the
end of the tip-ejector until it clicks
into place (see opposite).
To remove a tip-ejector extension (see
opposite):
- hold the pipette in one hand and
grip the extension with the other,
- gently twist the extension (either
direction) and pull it away from the
pipette.
Stroke Adjustment Wheel
Tip-ejector Button
(set for right handed user)
Slot
13