Hide thumbs Also See for PIPETMAN Classic P2:

Advertisement

CLASSIC
Quick Guide
EN

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PIPETMAN Classic P2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Gilson PIPETMAN Classic P2

  • Page 1 CLASSIC Quick Guide...
  • Page 2 Introduction PIPETMAN® Classic is a fully adjustable, air-displacement pipette that is used with disposable tips. Eight single channel models cover a volume range from 0.2 µl to 10 ml for many applications. Parts Checklist ● PIPETMAN Classic ● PIPETMAN Classic Quick Guide ●...
  • Page 3 These values are guaranteed only when genuine PIPETMAN® DIAMOND Tips are used. Each pipette is inspected and validated by qualified technicians in accordance with the Gilson Quality system. Gilson declares that its manufactured pipettes comply with the requirements of the IsO 8655 standard by type testing.
  • Page 4 Maximum Permissible Errors The data provided in the following tables conform to the IsO 8655-2 standard. GILSON ISO 8655 VOlUME (µl) SYSTEMATIC RANDOM SYSTEMATIC RANDOM ERROR (µL) ERROR (µL) ERROR (µL) ERROR (µL) P2 (P/N F144801) WITH D10 OR DL10 TIPS ≤ 0.012 ≤...
  • Page 5 Setting the Volume The volume of liquid to be aspirated is set using the volumeter. The dials are colored either black or red to indicate the position of the decimal point, depending on the model (refer to Figure 3). The volume is set by turning the thumbwheel or the push button.
  • Page 6 Pre-Rinse the Tips some liquids (e.g., protein-containing solutions and organic solvents) can leave a film of liquid on the inside wall of the tip. Pre-rinse the tip to minimize any errors that may be related to this phenomenon. Pre-rinsing consists of aspirating the first volume of liquid and then dispensing it back to waste.
  • Page 7 Gilson or its designates. Gilson’s sole liability under this warranty shall be limited to, at Gilson’s sole option, repair or replacement of any defective components of pipettes or refund of the purchase price paid for such pipettes.