PIPETMAN M Connected is a Bluetooth®-enabled motorized member of the ● Safety bag ● PIPETMAN M Connected pipette PIPETMAN® family of pipettes from Gilson, requiring virtually no pipetting forces to aspirate and dispense samples. Combining PIPETMAN’s renowned accuracy, ● Certificate of Conformity with ●...
Chapter 5 Chapter 4 Chapter 5 dESCriPTiON PiPETTiNG JUST LikE WiTh PiPETMAN Upper part Using PIPETMAN M Connected is as easy as using a mechanical Push button and volume PIPETMAN. Your PIPETMAN M Connected is set by default to “PiPET” adjustment knob Mode, which means forward pipetting.
Fit a tip suitable for the model Fit a tip suitable for the model of PIPETMAN M Connected that you are using of PIPETMAN M Connected (preferably use PIPETMAN Tips for assurance of accurate and precise results, refer to that you are using (preferably PiPETMAN diAMONd TiPS Chapter 9 on page 15).
REVERSE Mode Set the MiX volume: click the push button and set the volume by turning the push button. Confirm your settings with a simple click. This is the volume that will be The REVERSE Mode is reverse pipetting with a manual pipette. During aspirated and dispensed repeatedly.
To install PiPETMAN M Connected Software on a PC: data in the cloud for you to recall. It is compatible with the Gilson Connect platform of apps and has the ability to be connected to a Bluetooth 4.0 smart-ready tab or Download the PIPETMAN M Connected Software zip file and extract the installation phone or any PC equiped with Bluetooth.
Figure 6 identification number (ID). To ensure accuracy and precision, ID number Gilson’s quality assurance system focuses on the following critical parameters: If you want to connect to another pipette within the same channel as the first connected pipette, it will get disconnected and the new one will get connected while ●...
PiPETMAN Filter Tips as the filter will be damaged. Single Channel Model To ensure the best performance from your Gilson pipette, you should For single channel models, push the tip holder into the tip ALWAYS use PiPETMAN diAMONd Tips (in accordance with iSO8655)
Remove the window by inserting a small screwdriver ○ Never lay the pipette on its side when there is liquid in the tip(s) in the access slot. The Gilson Stand Adapter (refer to Parts Checklist and Accessories on page 5) Clip tag Position the name tag on the pipette.
USB port designed to ensure 900 pipetting cycles (750 for PIPETMAN M Connected Multichannel Use only the original power transformer supplied by Gilson. Use of an Plug 200 µL and 300 µL) at maximum speed (up incompatible power supply can damage your PiPETMAN M Connected! NOTICE to nine 96–well plates) between two charge...
○ Date accuracy, precision and robustness please proceed periodically with a two-minute ○ On/Off inspection as recommended by Gilson. Your PIPETMAN M Connected allows you to: NOTE Changes of service settings will reset all the service info to zero. ●...
These parts must be changed if they are damaged. You may also In accordance with ISO 8655 Gilson recommends a gravimetric procedure for remove these parts for cleaning or decontamination purposes. pipette calibration. This gravimetric method is used to establish the mean mass...
Lubricate the Piston (except P10M) Remove the tip ejector and (optionally) Only use Gilson lubricant (P/N: F2070902, as supplied). the lower part of the tip holder. If you Squeeze a small quantity from the tube onto a clean, lintless cloth. Use the cloth to remove the lower part, take care to transfer the lubricant to the piston.
Procedures for Gilson Pipettes, ejector locks, and ejector spacer. which you can download from Gilson’s website. Remove the tip ejector and the ejector spacer (refer to Chapter 15 MAiNTENANCE If you use chemical solutions for decontamination or detergents for cleaning, other page 22).
PIPETMAN DIAMOND Tips. The water level inside the tip should remain constant; if the level goes down then there is a Each pipette is inspected and validated by qualified technicians according to the Gilson leak (refer to Chapter 18...
DFL10ST ± 0.09 ≤ 0.06 ± 0.4 ≤ 0.2 1–20 µL 1–20 µL and spare parts. If the equipment is used in a manner not specified by Gilson in D200 DF30T ± 0.15 ≤ 0.10 ± 0.4 ≤ 0.2 P12x20M F81051 ±...
Chapter 21 APPENdiCES rEPLACEMENT PArTS Appendix A – Example of a Performance Check Single Channel Models Below is an example of how to evaluate the performance of PIPETMAN M Connected P10M at 1 µL. determine the mean value e Calculate the mean volume. DESCRIPTION P10M P20M...
Appendix B – Z Factor Appendix C – Evaporation Loss The reference calculation equation is : Procedure for the determination of Evaporation Loss z = [1/(P )] [1-(P Where: = density of air at t°C. Use the same distilled water, weighing vessel and balance as you will be using for the = density of the test liquid at t°C.
This warranty shall also be void in the event pipettes are 2014/53/EU Radio Equipment Directive altered or modified by any party other than Gilson or its designates. Gilson’s sole liability under this warranty shall be limited to, at Gilson’s sole option, repair or Standards for UK replacement of any defective components of pipettes or refund of the purchase price paid for such pipettes.