  • Page 1 User’s Guide...
  • Page 2: Table Of Contents

    BENEFIT FROM THE BLUETOOTH CONNECTION SAFETY PRECAUTIONS AND LIMITATIONS OF USE Tranfer Protocols on PC via Bluetooth | 13 Connect with Gilson Applications | 13 REPLACEMENT PARTS Connection with the PIPETMAN M Connected Software | 14 Single Channel Models | 34...
  • Page 3: Introduction

    PIPETMAN M Connected is a Bluetooth®-enabled motorized member of the ● Safety bag ● PIPETMAN M Connected pipette PIPETMAN® family of pipettes from Gilson, requiring virtually no pipetting forces to aspirate and dispense samples. Combining PIPETMAN’s renowned accuracy, ● Certificate of Conformity with ●...
  • Page 4: Description

    Chapter 5 Chapter 4 Chapter 5 dESCriPTiON PiPETTiNG JUST LikE WiTh PiPETMAN Upper part Using PIPETMAN M Connected is as easy as using a mechanical Push button and volume PIPETMAN. Your PIPETMAN M Connected is set by default to “PiPET” adjustment knob Mode, which means forward pipetting.
  • Page 5: Repetitive Mode

    Fit a tip suitable for the model Fit a tip suitable for the model of PIPETMAN M Connected that you are using of PIPETMAN M Connected (preferably use PIPETMAN Tips for assurance of accurate and precise results, refer to that you are using (preferably PiPETMAN diAMONd TiPS Chapter 9 on page 15).
  • Page 6: Reverse Mode

    REVERSE Mode Set the MiX volume: click the push button and set the volume by turning the push button. Confirm your settings with a simple click. This is the volume that will be The REVERSE Mode is reverse pipetting with a manual pipette. During aspirated and dispensed repeatedly.
  • Page 7: Custom Mode

    To install PiPETMAN M Connected Software on a PC: data in the cloud for you to recall. It is compatible with the Gilson Connect platform of apps and has the ability to be connected to a Bluetooth 4.0 smart-ready tab or Download the PIPETMAN M Connected Software zip file and extract the installation phone or any PC equiped with Bluetooth.
  • Page 8: Connection With The Pipetman M Connected Software

    Figure 6 identification number (ID). To ensure accuracy and precision, ID number Gilson’s quality assurance system focuses on the following critical parameters: If you want to connect to another pipette within the same channel as the first connected pipette, it will get disconnected and the new one will get connected while ●...
  • Page 9: Guidelines For Good Pipetting

    PiPETMAN Filter Tips as the filter will be damaged. Single Channel Model To ensure the best performance from your Gilson pipette, you should For single channel models, push the tip holder into the tip ALWAYS use PiPETMAN diAMONd Tips (in accordance with iSO8655)
  • Page 10: General Guidelines For Good Pipetting

    Remove the window by inserting a small screwdriver ○ Never lay the pipette on its side when there is liquid in the tip(s) in the access slot. The Gilson Stand Adapter (refer to Parts Checklist and Accessories on page 5) Clip tag Position the name tag on the pipette.
  • Page 11: Alarms

    USB port designed to ensure 900 pipetting cycles (750 for PIPETMAN M Connected Multichannel Use only the original power transformer supplied by Gilson. Use of an Plug 200 µL and 300 µL) at maximum speed (up incompatible power supply can damage your PiPETMAN M Connected! NOTICE to nine 96–well plates) between two charge...
  • Page 12: Configuration

    ○ Date accuracy, precision and robustness please proceed periodically with a two-minute ○ On/Off inspection as recommended by Gilson. Your PIPETMAN M Connected allows you to: NOTE Changes of service settings will reset all the service info to zero. ●...
  • Page 13: Adjustment

    These parts must be changed if they are damaged. You may also In accordance with ISO 8655 Gilson recommends a gravimetric procedure for remove these parts for cleaning or decontamination purposes. pipette calibration. This gravimetric method is used to establish the mean mass...
  • Page 14: Maintenance Operation For Multichannel Models

    Lubricate the Piston (except P10M) Remove the tip ejector and (optionally) Only use Gilson lubricant (P/N: F2070902, as supplied). the lower part of the tip holder. If you Squeeze a small quantity from the tube onto a clean, lintless cloth. Use the cloth to remove the lower part, take care to transfer the lubricant to the piston.
  • Page 15: Cleaning And Decontamination

    Procedures for Gilson Pipettes, ejector locks, and ejector spacer. which you can download from Gilson’s website. Remove the tip ejector and the ejector spacer (refer to Chapter 15 MAiNTENANCE If you use chemical solutions for decontamination or detergents for cleaning, other page 22).
  • Page 16: Leak Test

    PIPETMAN DIAMOND Tips. The water level inside the tip should remain constant; if the level goes down then there is a Each pipette is inspected and validated by qualified technicians according to the Gilson leak (refer to Chapter 18...
  • Page 17: Safety Precautions And Limitations Of Use

    DFL10ST ± 0.09 ≤ 0.06 ± 0.4 ≤ 0.2 1–20 µL 1–20 µL and spare parts. If the equipment is used in a manner not specified by Gilson in D200 DF30T ± 0.15 ≤ 0.10 ± 0.4 ≤ 0.2 P12x20M F81051 ±...
  • Page 18: Replacement Parts

    Chapter 21 APPENdiCES rEPLACEMENT PArTS Appendix A – Example of a Performance Check Single Channel Models Below is an example of how to evaluate the performance of PIPETMAN M Connected P10M at 1 µL. determine the mean value e Calculate the mean volume. DESCRIPTION P10M P20M...
  • Page 19: Appendix B - Z Factor

    Appendix B – Z Factor Appendix C – Evaporation Loss The reference calculation equation is : Procedure for the determination of Evaporation Loss z = [1/(P )] [1-(P Where: = density of air at t°C. Use the same distilled water, weighing vessel and balance as you will be using for the = density of the test liquid at t°C.
  • Page 20: Regulatory Compliance

    This warranty shall also be void in the event pipettes are 2014/53/EU Radio Equipment Directive altered or modified by any party other than Gilson or its designates. Gilson’s sole liability under this warranty shall be limited to, at Gilson’s sole option, repair or Standards for UK replacement of any defective components of pipettes or refund of the purchase price paid for such pipettes.
  • Page 21 LT801594/G | ©2021 Gilson, Inc. All rights reserved.