Gilson pipetman P2G Manual
Hide thumbs Also See for pipetman P2G:

Advertisement

Quick Links

Trusted results
Best results guaranteed with a performance exceeding ISO recommendations.
Maximum Permissible Errors
Volume
Gilson
Model
tested
Systematic error
Random error
(μL)
(μL)
P2G
Min.:
0.2
± 0.024
≤ 0.012
(F144054P)
0.5
± 0.025
≤ 0.012
POS. 1 066803
(F144054M)
Max.:
2
± 0.030
≤ 0.014
P10G
Min.:
1
± 0.025
≤ 0.012
(F144055P)
5
± 0.075
≤ 0.030
(F144055M)
Max.:
10
± 0.100
≤ 0.040
POS. 2 066805
P20G
Min.:
2
± 0.10
≤ 0.030
(F144056P)
10
± 0.10
≤ 0.050
POS. 3 066807
(F144056M)
Max.:
20
± 0.20
≤ 0.060
P100G
Min.:
10
± 0.35
≤ 0.10
(F144057P)
50
± 0.40
≤ 0.12
(F144057M)
Max.:
100
± 0.80
≤ 0.15
P200G
Min.:
20
± 0.50
≤ 0.20
(F144058P)
100
± 0.80
≤ 0.25
(F144058M)
Max.:
200
± 1.60
≤ 0.30
POS. 4 066811
P1000G
Min.:
100
± 3.0
≤ 0.6
(F144059P)
500
± 4.0
≤ 1.0
Max.:
1000
± 8.0
≤ 1.5
POS. 5 066813
(F144059M)
P5000G
Min.:
500
± 12
(F144066)
2500
± 15
Max.:
5000
± 30
P10mL G
Min.:
1000
± 30
(F144067)
5000
± 40
≤ 10
Max.:
10000
± 60
≤ 16
Pipette Kits
All you need to get started, in a single box!
PIPETMAN G kits complement the product range and offer
cost-effective package to get you started or to renew your
old set of pipette.
PIPETMAN G
Ordering reference
Pipette Models*
STARTER KIT
F167900
P20G - P200G - P1000G
MICRO-VOLUME KIT
F167800
P2G - P10G - P100G
*With both ejectors: stainless steel and plastic ejectors.
ISO 8655
Systematic error
Random error
(μL)
(μL)
± 0.08
≤ 0.04
± 0.08
≤ 0.04
± 0.08
≤ 0.04
± 0.12
≤ 0.08
± 0.12
≤ 0.08
± 0.12
≤ 0.08
± 0.2
≤ 0.1
± 0.2
≤ 0.1
± 0.2
≤ 0.1
Pipette model
± 0.8
≤ 0.3
± 0.8
≤ 0.3
± 0.8
≤ 0.3
± 1.6
≤ 0.6
± 1.6
≤ 0.6
± 1.6
≤ 0.6
± 8.0
≤ 3.0
± 8.0
≤ 3.0
± 8.0
≤ 3.0
≤ 3
± 40
≤ 15
≤ 5
± 40
≤ 15
≤ 8
± 40
≤ 15
≤ 6
± 60
≤ 30
± 60
≤ 30
± 60
≤ 30
PIPETMAN DIAMOND Tips
D200 - D1000
DFL10ST - DF100ST
w w w. g il s on.c om I
Gilson, Inc. World Headquarters I
Middleton, WI 53562-0027 USA
Tel: (1) 800-445-7661 or (1) 608-836-1551
Gilson S.A.S. I
19, avenue des Entrepreneurs
I
BP 145, F-95400 Villiers-le-Bel, FRANCE
Tel: (33-1) 34 29 50 00
Gilson's Complete Pipetting System
Your protocols and experiments require more than just generic pipette tips. Indeed, tips
are essential components of your pipetting system. According to ISO 8655-2*:
"The maximum permissible errors always apply to the total system of piston pipette
and tip. When using pipette tips not delivered by the pipette supplier, the supplier
declaration or certificate of conformity does not apply".
*ISO 8655 is an international standard for "piston operated apparatus". It defines, among other things, terms and
definitions regarding piston operated pipettes and their testing, testing procedure and maximum permissible errors.
Recommended PIPETMAN DIAMOND tips
Model (scale 1/2)
Volume range
D10
DL10
P2G
DF10
DFL10
D10
DL10
P10G
DF10
DFL10
D200
P20G
DF30
D200
P100G
DF100
D200
D300
P200G
DF200
DF300
D1000
D1200
P1000G
DF1000
DF1200
P5000 G
D5000
P10mL G
D10mL
sa l e s@ g i lso n. co m
I
ser vi ce @ gi l so n .co m
I
tra i ni n g @gi lso n.com
3000 Parmenter Street I P.O. Box 620027
I
Fax: (1) 608-831-4451
I
Fax: (33-1) 34 29 50 20
0.1 – 10 µL
0.1 – 20 µL
0.1 – 10 µL
0.1 – 10 µL
0.1 – 10 µL
0.1 – 20 µL
0.1 – 10 µL
0.1 – 10 µL
2 – 200 µL
2 – 30 µL
2 – 200 µL
20 – 100 µL
2 – 200 µL
20 – 300 µL
20 – 200 µL
20 – 300 µL
100 – 1000 µL
100 – 1200 µL
100 – 1000 µL
100 – 1200 µL
1 – 5 mL
1 – 10 mL
LT800511/C © 05/2012 Gilson SAS.
All rights reserved. All trademarks
are the property of Gilson Inc. and
its subsidiaries.
Premium pipetting comfort!
EXTREMELY LOW PIPETTING FORCEs!
EXTREMELY EAsY TIP EJECTION!
LEGENdARY PIPETMAN RObusTNEss!

Advertisement

Table of Contents

Troubleshooting

loading

Summary of Contents for Gilson pipetman P2G

  • Page 1 19, avenue des Entrepreneurs LT800511/C © 05/2012 Gilson SAS. All rights reserved. All trademarks BP 145, F-95400 Villiers-le-Bel, FRANCE are the property of Gilson Inc. and Tel: (33-1) 34 29 50 00 Fax: (33-1) 34 29 50 20 its subsidiaries.
  • Page 2 Enjoy a new addition to PIPETMAN line with its legendary accuracy, precision, and *P5000G and P10mL G models do not require tip ejectors. **PIPETMAN G can only be lubricated with a high quality lubricant selected and approved by Gilson. The use of other types of lubricants robustness… with the highest comfort! will void the warranty for your pipette.
  • Page 3 • Progettate, prodotte, calibrate e validate in conformità con le specifiche e i metodi dello standard ISO8655. • Corredate di certificato di conformità originale Gilson. • Conformi alla Normativa IVD (98/79/EC) per gli strumenti In – Vitro Diagnostici. • Garanzia 12 mesi.
  • Page 4 Premium Pipetting Comfort! IMPORTANT NOTICE! This pipette can be only lubricated with a high quality lubricant selected and approved by Gilson. This lubricant is available under the reference: 5440011070. The use of other type of lubricant cancels the warranty of this pipette.
  • Page 5: Table Of Contents

    PIPETMAN G is a fully adjustable air displacement pipette and is used with disposable tips. To answer the current needs of our loyal users of ours pipettes, and still being conform to Gilson brand qualities, three improvements have been added to our mechanical series: ...
  • Page 6: Parts Check List

    The use of other lubricant cancels the warranty of this pipette. This lubricant is available under the reference: 5440011070. Please contact your Gilson distributor. 2 - Parts check LIst Just take a moment to verify that the following items are present: ...
  • Page 7: Setting The Volume

    4 - settIng the VoLume The volume of liquid to be aspi- P10G P20G rated is set using the volumeter. The dials are colored either black or red to indicate the position of the decimal point, 1.25 µL 7.5 µL 12.5 µL depending on the model (see P100G...
  • Page 8 Gilson logo engraved on their collar, ensuring that you have a genuine Gilson product. Plastic tips are for a single application – they must not be cleaned for reuse. PIPetman G can also be used with the main tip brands.
  • Page 9 Pre-rinse the tips Some liquids (e.g. protein-containing solutions and organic solvents) can leave a film of liquid on the inside wall of the tip ; pre-rinse the tip to minimize any errors that may be related to this phenomenon. Pre-rinsing consists of aspirating the first volume of liquid and then dispensing it back into the same vessel (or to waste).
  • Page 10: General Guidelines For Good Pipetting

    dispense Place the end of the tip against the inside wall of the recipient vessel (at an angle of 10° to 40°). Press the push-button slowly and smoothly to the first stop. Wait for at least a second, then press the push-button to the second stop to expel any residual liquid from the tip.
  • Page 11: Accessories

    7 - accessorIes To make pipetting more comfortable and more secure, Gilson has developed several accessories: To avoid the possibility of liquid running back into the pipette, store the pipette vertically. Ca RROUseL™ Pipette stand (7 pipettes) f 161401 TRIO™...
  • Page 12: Glp Features

    Before returning any pipette to your local Gilson Service Center, ensure that it is completely free of chemical, biological, or radioactive contamination. Refer to Chapter 12 - Cleaning and Decontamination. Please use the included safety bag to return the pipette to your local Gilson Service Center.
  • Page 13 Replace the ejector spacer The tip-ejector is loose assemble the tip-ejector properly The ejector lock is misaligned align the ejector lock Pipetting seize up Piston need lubricant Lubricate piston assembly However, if you can’t solve the problem, contact your Gilson distributor.
  • Page 14: Leak Test

    10 - Leak test This test may be performed at any time to check that the pipette does not leak, especially after performing a main- tenance or decontamination procedure. If a pipette fails this test, replace the O-ring and seal. After making sure that the pipette is correctly reassembled, repeat this test.
  • Page 15: Maintenance

    Replacing spare parts  Greasing the piston assembly. PIPeTmaN P2G and P10G should not be disassembled, so you may only replace the push-button, tip-ejector, dual position tip-ejector and its adapter. With these pipettes if the tip-holder is damaged, the piston may also be damaged.
  • Page 16 Color coded push-button Blue ring Connecting nut P10G, P20G P5000G P100G, P200G P1000G P10mlG Piston assembly Tip-holder Tip-ejector stainless steel plastic...
  • Page 17 You may remove the piston assembly for cleaning pur- poses only. If the piston assembly is changed, the pipette must be adjusted and calibrated in a Gilson authorized Service Center. As the models P2G and P10G contain miniaturized parts, it is best not to disassemble these pipettes yourself.
  • Page 18: Cleaning And Decontamination

    As the models P2G and P10G contain miniaturized parts, it is best not to disassemble these pipettes yourself, please contact your local Gilson authorized Service Center. The dimensions of the O-ring vary depending on the model of pipette.
  • Page 19 You may refer to the decontamination procedure available on the Gilson website (www. gilson.com). Liquid must never enter the upper part (handle) of any pipette. cleaning The pipette must be cleaned, as described below, before it is decontaminated. Soap solution is recommended for cleaning PIPETMAN G.
  • Page 20 Reassemble the pipette as described in the Chapter 11 - Maintenance. autoclaving The upper part (body) and the piston assembly of the pipette are not autoclavable. Only the following parts may be autoclaved: tip-ejector, tip-holder and connecting nut. The O-ring and seal are not autoclavable; they may be cleaned or replaced with the one specified in Chapter 14 - Spare Parts.
  • Page 21: Specifications

    figures are guaranteed only when genuine PIPETMAN DIAMOND Tips are used. Each pipette is inspected and validated by qualified technicians in accordance with the Gilson Quality System. Gilson declares that its manufactured pipettes comply with the requirements of the ISO 8655 standard, by type testing.
  • Page 22 Permissible errors Maximum Permissible Errors Model Volume Gilson ISO 8655 (Reference) (µL) systematic Random systematic Random error (µL) error (µL) error (µL) error (µL) ± 0.024  0.012 ± 0.08  0.04 (f 144054P) ± 0.025  0.012 ±...
  • Page 23: Spare Parts

    14 - sPare Parts Service Kit 1 level includes: - 3 piston seals or seal guides - 3 O-rings - 1 tip-holder Service Kit 2 level includes: - 1 push-button - 1 connecting nut - 2 tip-ejectors F1 F2 for the P2G and P10G models only: - 1 adapter for stainless steel tip-ejector...
  • Page 24 P2g (F144054P or m) and P10g (F144055P or m) Description P10G C+D+E F144501 FA07001 Service Kit 1st level A+B+F1 to F4 F1619702 F1619712 Service Kit 2nd level Seal + O-ring (5 sets) F144861 FA07012 F144879 F144879 Tip-ejector adapter F2070903 F2070903 Tip-ejector extension P20g (F144056 P or m) and P100g (F144057 P or m) Description...
  • Page 25: Ec Declaration Of Conformity

    The company, gILson s.a.s. 19, avenue des Entrepreneurs BP. 145 - 95400 Villiers-le-Bel (France) Tel.: +33(0) 1 34 29 50 00 Fax: +33(0) 1 34 29 50 20 www.gilson.com Hereby certifies on its sole responsibility that the products listed below: PIPetman®...
  • Page 26: Warranty

    Gilson or its designates. Gilson’s sole liability under this warranty shall be limited to, at Gilson’s sole option, repair or replacement of any defective components of pipettes or refund of the purchase price paid for such pipettes.
  • Page 27 LT801122/B - ©2012 Gilson SAS All rights reserved May 2012 English Printed in France Specifications subject to change without notifications - errors omitted. www.gilson.com sales@gilson.com service@gilson.com training@gilson.com gilson, Inc. World headquarters 3000 Parmenter Street P.O. Box 620027 Middleton, WI 53562-0027, USA...
  • Page 28 Premium Pipetting Comfort! IMPORTANT NOTICE! This pipette can be only lubricated with a high quality lubricant selected and approved by Gilson. This lubricant is available under the reference: 5440011070. The use of other type of lubricant cancels the warranty of this pipette.
  • Page 29 PIPETMAN G is a fully adjustable air displacement pipette and is used with disposable tips. To answer the current needs of our loyal users of ours pipettes, and still being conform to Gilson brand qualities, three improvements have been added to our mechanical series: ...
  • Page 30: Parts Check List

    The use of other lubricant cancels the warranty of this pipette. This lubricant is available under the reference: 5440011070. Please contact your Gilson distributor. 2 - Parts check LIst Just take a moment to verify that the following items are present: ...
  • Page 31: Setting The Volume

    4 - settIng the VoLume The volume of liquid to be aspi- P10G P20G rated is set using the volumeter. The dials are colored either black or red to indicate the position of the decimal point, 1.25 µL 7.5 µL 12.5 µL depending on the model (see P100G...
  • Page 32 Gilson logo engraved on their collar, ensuring that you have a genuine Gilson product. Plastic tips are for a single application – they must not be cleaned for reuse. PIPetman G can also be used with the main tip brands.
  • Page 33 Pre-rinse the tips Some liquids (e.g. protein-containing solutions and organic solvents) can leave a film of liquid on the inside wall of the tip ; pre-rinse the tip to minimize any errors that may be related to this phenomenon. Pre-rinsing consists of aspirating the first volume of liquid and then dispensing it back into the same vessel (or to waste).
  • Page 34: General Guidelines For Good Pipetting

    dispense Place the end of the tip against the inside wall of the recipient vessel (at an angle of 10° to 40°). Press the push-button slowly and smoothly to the first stop. Wait for at least a second, then press the push-button to the second stop to expel any residual liquid from the tip.
  • Page 35: Accessories

    7 - accessorIes To make pipetting more comfortable and more secure, Gilson has developed several accessories: To avoid the possibility of liquid running back into the pipette, store the pipette vertically. CaRROUseL™ Pipette stand (7 pipettes) f161401 TRIO™...
  • Page 36: Glp Features

    Before returning any pipette to your local Gilson Service Center, ensure that it is completely free of chemical, biological, or radioactive contamination. Refer to Chapter 12 - Cleaning and Decontamination. Please use the included safety bag to return the pipette to your local Gilson Service Center.
  • Page 37 Replace the ejector spacer The tip-ejector is loose assemble the tip-ejector properly The ejector lock is misaligned align the ejector lock Pipetting seize up Piston need lubricant Lubricate piston assembly However, if you can’t solve the problem, contact your Gilson distributor.
  • Page 38: Leak Test

    10 - Leak test This test may be performed at any time to check that the pipette does not leak, especially after performing a main- tenance or decontamination procedure. If a pipette fails this test, replace the O-ring and seal. After making sure that the pipette is correctly reassembled, repeat this test.
  • Page 39: Maintenance

    Replacing spare parts  Greasing the piston assembly. PIPeTmaN P2G and P10G should not be disassembled, so you may only replace the push-button, tip-ejector, dual position tip-ejector and its adapter. With these pipettes if the tip-holder is damaged, the piston may also be damaged.
  • Page 40 Color coded push-button Blue ring Connecting nut P10G, P20G P5000G P1000G P100G, P200G P10mlG Piston assembly Tip-holder Tip-ejector stainless steel plastic...
  • Page 41 You may remove the piston assembly for cleaning pur- poses only. If the piston assembly is changed, the pipette must be adjusted and calibrated in a Gilson authorized Service Center. As the models P2G and P10G contain miniaturized parts, it is best not to disassemble these pipettes yourself.
  • Page 42: Cleaning And Decontamination

    As the models P2G and P10G contain miniaturized parts, it is best not to disassemble these pipettes yourself, please contact your local Gilson authorized Service Center. The dimensions of the O-ring vary depending on the model of pipette.
  • Page 43 You may refer to the decontamination procedure available on the Gilson website (www. gilson.com). Liquid must never enter the upper part (handle) of any pipette. cleaning The pipette must be cleaned, as described below, before it is decontaminated. Soap solution is recommended for cleaning PIPETMAN G.
  • Page 44 Reassemble the pipette as described in the Chapter 11 - Maintenance. autoclaving The upper part (body) and the piston assembly of the pipette are not autoclavable. Only the following parts may be autoclaved: tip-ejector, tip-holder and connecting nut. The O-ring and seal are not autoclavable; they may be cleaned or replaced with the one specified in Chapter 14 - Spare Parts.
  • Page 45: Specifications

    figures are guaranteed only when genuine PIPETMAN DIAMOND Tips are used. Each pipette is inspected and validated by qualified technicians in accordance with the Gilson Quality System. Gilson declares that its manufactured pipettes comply with the requirements of the ISO 8655 standard, by type testing.
  • Page 46 Permissible errors Maximum Permissible Errors Model Volume Gilson ISO 8655 (Reference) (µL) systematic Random systematic Random error (µL) error (µL) error (µL) error (µL) ± 0.024  0.012 ± 0.08  0.04 (f144054P) ± 0.025  0.012 ± 0.08 ...
  • Page 47: Spare Parts

    14 - sPare Parts Service Kit 1 level includes: - 3 piston seals or seal guides - 3 O-rings - 1 tip-holder Service Kit 2 level includes: - 1 push-button - 1 connecting nut - 2 tip-ejectors F1 F2 for the P2G and P10G models only: - 1 adapter for stainless steel tip-ejector...
  • Page 48 P2g (F144054P or m) and P10g (F144055P or m) Description P10G C+D+E Service Kit 1st level F144501 FA07001 Service Kit 2nd level A+B+F1 to F4 F1619702 F1619712 Seal + O-ring (5 sets) F144861 FA07012 Tip-ejector adapter F144879 F144879 Tip-ejector extension F2070903 F2070903 P20g (F144056 P or m) and P100g (F144057 P or m)
  • Page 49: Ec Declaration Of Conformity

    The company, gILson s.a.s. 19, avenue des Entrepreneurs BP. 145 - 95400 Villiers-le-Bel (France) Tel.: +33(0) 1 34 29 50 00 Fax: +33(0) 1 34 29 50 20 www.gilson.com Hereby certifies on its sole responsibility that the products listed below: PIPetman®...
  • Page 50 notes...
  • Page 51 LT801122/A - ©2011 Gilson SAS All rights reserved March 2011 English Printed in France Specifications subject to change without notifications - errors omitted. www.gilson.com sales@gilson.com service@gilson.com training@gilson.com gilson, Inc. World headquarters 3000 Parmenter Street P.O. Box 620027 Middleton, WI 53562-0027, USA...