Trusted results
Best results guaranteed with a performance exceeding ISO recommendations.
Maximum Permissible Errors
Volume
Gilson
Model
tested
Systematic error
Random error
(μL)
(μL)
P2G
Min.:
0.2
± 0.024
≤ 0.012
(F144054P)
0.5
± 0.025
≤ 0.012
POS. 1 066803
(F144054M)
Max.:
2
± 0.030
≤ 0.014
P10G
Min.:
1
± 0.025
≤ 0.012
(F144055P)
5
± 0.075
≤ 0.030
(F144055M)
Max.:
10
± 0.100
≤ 0.040
POS. 2 066805
P20G
Min.:
2
± 0.10
≤ 0.030
(F144056P)
10
± 0.10
≤ 0.050
POS. 3 066807
(F144056M)
Max.:
20
± 0.20
≤ 0.060
P100G
Min.:
10
± 0.35
≤ 0.10
(F144057P)
50
± 0.40
≤ 0.12
(F144057M)
Max.:
100
± 0.80
≤ 0.15
P200G
Min.:
20
± 0.50
≤ 0.20
(F144058P)
100
± 0.80
≤ 0.25
(F144058M)
Max.:
200
± 1.60
≤ 0.30
POS. 4 066811
P1000G
Min.:
100
± 3.0
≤ 0.6
(F144059P)
500
± 4.0
≤ 1.0
Max.:
1000
± 8.0
≤ 1.5
POS. 5 066813
(F144059M)
P5000G
Min.:
500
± 12
(F144066)
2500
± 15
Max.:
5000
± 30
P10mL G
Min.:
1000
± 30
(F144067)
5000
± 40
≤ 10
Max.:
10000
± 60
≤ 16
Pipette Kits
All you need to get started, in a single box!
PIPETMAN G kits complement the product range and offer
cost-effective package to get you started or to renew your
old set of pipette.
PIPETMAN G
Ordering reference
Pipette Models*
STARTER KIT
F167900
P20G - P200G - P1000G
MICRO-VOLUME KIT
F167800
P2G - P10G - P100G
*With both ejectors: stainless steel and plastic ejectors.
ISO 8655
Systematic error
Random error
(μL)
(μL)
± 0.08
≤ 0.04
± 0.08
≤ 0.04
± 0.08
≤ 0.04
± 0.12
≤ 0.08
± 0.12
≤ 0.08
± 0.12
≤ 0.08
± 0.2
≤ 0.1
± 0.2
≤ 0.1
± 0.2
≤ 0.1
Pipette model
± 0.8
≤ 0.3
± 0.8
≤ 0.3
± 0.8
≤ 0.3
± 1.6
≤ 0.6
± 1.6
≤ 0.6
± 1.6
≤ 0.6
± 8.0
≤ 3.0
± 8.0
≤ 3.0
± 8.0
≤ 3.0
≤ 3
± 40
≤ 15
≤ 5
± 40
≤ 15
≤ 8
± 40
≤ 15
≤ 6
± 60
≤ 30
± 60
≤ 30
± 60
≤ 30
PIPETMAN DIAMOND Tips
D200 - D1000
DFL10ST - DF100ST
w w w. g il s on.c om I
Gilson, Inc. World Headquarters I
Middleton, WI 53562-0027 USA
Tel: (1) 800-445-7661 or (1) 608-836-1551
Gilson S.A.S. I
19, avenue des Entrepreneurs
I
BP 145, F-95400 Villiers-le-Bel, FRANCE
Tel: (33-1) 34 29 50 00
Gilson's Complete Pipetting System
Your protocols and experiments require more than just generic pipette tips. Indeed, tips
are essential components of your pipetting system. According to ISO 8655-2*:
"The maximum permissible errors always apply to the total system of piston pipette
and tip. When using pipette tips not delivered by the pipette supplier, the supplier
declaration or certificate of conformity does not apply".
*ISO 8655 is an international standard for "piston operated apparatus". It defines, among other things, terms and
definitions regarding piston operated pipettes and their testing, testing procedure and maximum permissible errors.
Recommended PIPETMAN DIAMOND tips
Model (scale 1/2)
Volume range
D10
DL10
P2G
DF10
DFL10
D10
DL10
P10G
DF10
DFL10
D200
P20G
DF30
D200
P100G
DF100
D200
D300
P200G
DF200
DF300
D1000
D1200
P1000G
DF1000
DF1200
P5000 G
D5000
P10mL G
D10mL
sa l e s@ g i lso n. co m
I
ser vi ce @ gi l so n .co m
I
tra i ni n g @gi lso n.com
3000 Parmenter Street I P.O. Box 620027
I
Fax: (1) 608-831-4451
I
Fax: (33-1) 34 29 50 20
0.1 – 10 µL
0.1 – 20 µL
0.1 – 10 µL
0.1 – 10 µL
0.1 – 10 µL
0.1 – 20 µL
0.1 – 10 µL
0.1 – 10 µL
2 – 200 µL
2 – 30 µL
2 – 200 µL
20 – 100 µL
2 – 200 µL
20 – 300 µL
20 – 200 µL
20 – 300 µL
100 – 1000 µL
100 – 1200 µL
100 – 1000 µL
100 – 1200 µL
1 – 5 mL
1 – 10 mL
LT800511/C © 05/2012 Gilson SAS.
All rights reserved. All trademarks
are the property of Gilson Inc. and
its subsidiaries.
Premium pipetting comfort!
EXTREMELY LOW PIPETTING FORCEs!
EXTREMELY EAsY TIP EJECTION!
LEGENdARY PIPETMAN RObusTNEss!