5. Communication

5.1. Device Address

Sets the device address or poll ID on a HART network. Any address other than 0 will
cause the output current to be a fixed value, and the current will not indicate the
reading.
Values
To set the address via the handheld programmer:
Press Right ARROW
Press Down ARROW
item 5).
Press Right ARROW
Scroll to DEVICE ADDRESS and press Right ARROW
view.
Press Right ARROW
ARROW

5.2. Communication Control

SITRANS LR250 can only reset this parameter via the handheld
Note:
programmer.
Enables /disables the read/write access to parameters via remote communications.
Options
Page 86
SITRANS LR250 (HART) – INSTRUCTION MANUAL
Range: 0 to 15
Default: 0
to activate PROGRAM mode and open menu level 1.
repeatedly to navigate to COMMUNICATION (menu
to display the parameter list.
to open Edit mode. Key in a new value and press Right
to accept it.
No changes are permitted via remote
Read Only
communications.
*
Read Write
Changes are permitted.
Sets the status to Read Only, with the potential
Restricted
for another HART device to change this via
remote communications.
to open parameter
7ML19985JE01