8) The downgrade
process
starts
and
"Upgrading..."
appears on the
bottom
of
window (but the
firmware is really
downgraded).
9) Close the "firmware downgrading/upgrading" window and EC-Lab V10.23. The
downgrading procedure is now completed and you can connect the older version to the
instrument.
10) Open the older version of EC-Lab
40
the
Fig. 53: Upgrading message on the status bar.
V10.12
Manual for VMP300-based instruments
Fig. 52: Flash loading.
and connect to the instrument.