Optional Accessories

6.2.5 Super mega-cassette clamp

Assembly of the super mega-cassette clamp
The super mega cassette clamp should preferably be used with the rigid fixture for specimen
clamps.
To do so, proceed as follows:
Remove the rubber ring (5) only after having attached the cassette clamp on the specimen
head!
• Screw the rigid fixture for specimen clamps (4) onto the specimen head (3):
Remove the screw (1), place the specimen holder fixture (4) onto the specimen head (3) from
the front and tighten the screws (2) with an Allen key No. 3.
Next, insert the screw (1) from the side and briefly tighten it with an Allen key No. 4.
• Insert the super mega cassette clamp from the side on the left into the dovetail guide of the
rigid fixture for specimen clamps and tighten screw (1).
If the directional fixture for specimen clamps is used with the rigid knife holder base, the
orientation must be in position "0" and the cover for backlighting illumination must be
detached. (Danger of collision if not observed!)
NEVER use the super mega-cassette clamp with backlighting illumination!