2. Attach vertical column
• Insert the cap screw (b) into the hole of the
baseplate from below. Place the lock washer
(c) on the cap screw from above.
• Thread the silver vertical column (2) onto the
baseplate from above and tighten with the
Allen key No. 8.
The lock washer must be positioned
between the baseplate and vertical
column to prevent unintentional turn-
ing of the column.
Extreme danger of tipping! After in-
stalling the vertical column, imme-
diately place the microtome onto the
baseplate so that the front feet of the
microtome are positioned in the shal-
low recesses (A).
a
3. Attach horizontal arm
• Slide the support ring (e) onto the vertical
column and position it so that the lock nut (f)
is facing to the back. Tighten the lock nut.
g
• Slide the cross-member (d) onto the column.
Ensure that the lock screw (g) is facing to the
right of the baseplate. The horizontal arm must
be centered over the microscope.
• Slide the horizontal arm (3), flat side facing the
lock screw (g), into the cross-member (d) and
tighten.
For more information on connecting
and using the microscope, magnifying
lens or cold-light source, please see
the appropriate Instructions for Use.