6.
Optional accessories

6.2.6 Super mega-cassette clamp

Assembly of the super mega-cassette clamp
36
2
The super mega-cassette clamp should preferably be used with the rigid
fixture for specimen clamps.
To do so, proceed as follows:
Remove the rubber ring (5) only after having attached the cassette
clamp on the object head!
• Screw the rigid fixture for specimen clamps (4) onto the object head (3):
Remove the screw (1), place the fixture for specimen clamps (4) onto
the object head (3) from the front and tighten the screws (2) with an
Allen key size 3.
Next, insert the screw (1) from the side and briefly tighten it with an
Allen key size 4.
• Insert the super mega-cassette clamp from the side on the left into the
dovetail guide of the rigid fixture for specimen clamps and tighten
screw (1).
If the directional fixture for specimen clamps is used with the rigid
knife holder base, the orientation must be in position "0" and the
cover for backlighting illumination must be detached. (Danger of
collision if not observed!)
NEVER use the super mega-cassette clamp with backlighting
illumination!
4
5
1
Instruction manual V 1.4 – 02/2010
3
Fig. 32