8.
Cleaning and Maintenance
95
96
91
56
11
88
97
Knife holder N
70
55
10
8
Knife holder NZ
73
Fig. 95
8.3

Lubricating the instrument

Once a month, lubricate the following parts with
the included drive parts oil No. 405 (1 - 2 drops
are enough).
Instrument and specimen holder:
• The clamping key (95) of the quick clamping
system.
• The T-piece (55) on the microtome base plate.
• Guide rails (96) for the knife holder base on the
microtome base plate.
• The T-piece on the knife holder base (56).
Knife holder E:
• Clamping lever (11) for the lateral displace-
ment.
• T-piece (88) and guide (97) of the knife holder
for the lateral displacement.
• Clamping lever (10) for the blade.
Knife holders N and NZ:
• Sliding surfaces of the finger guard (8).
• Knurled nuts (73) for measuring height adjust-
ment.