Step 5: Eliminate mounting effects
Devices are factory-calibrated. Once installed, it is recommended to perform
this step to eliminate potential error caused by mounting position or static
pressure. Instructions for using a Field Communicator are listed below.
Note
See the Rosemount Smart Pressure Gauge
Using AMS Configurator
Sensor trim function on absolute gauge
1. Vent the device.
2. Connect the Field Communicator.
3. From the HOME screen, enter the HART Fast Key sequence.
Device dashboard Fast Keys
4. Follow the commands to perform the procedure.
4.0

Troubleshooting

This section provides information for basic troubleshooting. See the
Manual
for advanced troubleshooting.
Device status
The flashing LED indicates device status using the colors descibed in
Table 1. Status Descriptions
LED color
Pressure measurement
If the mounting effects have not been eliminated after completing
perform this alternative procedure for verifying the pressure value.
1. From the HOME screen, eter the HART Fast Key sequence.
Device dashboard Fast Keys
2. Follow the commands to perform the procedure.
2, 1, 1
Green
Amber
Battery is low, battery replacement recommended
Red
No
No power, verify ON/OFF switch is in "on" position
color
2, 2, 1, 1, 1
Reference Manual
Device status
Functioning properly
Battery replacement required
OR
Device is malfunctioning
Quick Start Guide
for the following:
Reference
Table
Step
5,
1.
7