SP600 installation and user's guide
www.renishaw.com
Technical specification
Probe
Probe attributes
Measurement range
Overtravel range
Resolution with optional AC2 interface
card
Resolution with optional AC1 interface
card
Spring rate
Damping
Weight
Power supply
Outputs (X, Y, Z)
Mounting
Interface
Cabling
Outputs
Voltage output scaling
Axis
Stylus length 50 mm (1.968 in)
X
5.6 V/mm
Y
6.3 V/mm
Z
5 V/mm
The SP600 is designed to have 5 V/mm @ 100 mm stylus tip.
Issued 01 2018
3 axis measurement (X, Y, Z)
Linear and parallel motion in all axes
±1 mm (±0.04 in) in all axes and orientations with a 50 mm (1.97 in) 4.6 g (0.16 oz)
stylus
X, Y and -Z protected by a kinematic joint
0.1 μm (0.000004 in)
1 μm (0.00004 in)
120 gmf/mm nominal (X, Y, Z)
20% (X, Y, Z) typical
SP600 probe: 172 g (6.1 oz) (excluding shank)
SP600M probe: 216 g (7.6 oz)
SP600Q probe: 299 g (10.5 oz)
+12 V to 12 V, 5 V (±10%)
Analogue proportional
Voltage output scaling: 4 V to 8.5 V/mm (dependant on stylus)
SP600 adaptors to suit clutch, shank adaptor or CMM shank
SP600M autojoint
SP600Q is mounted directly to the quill of the CMM
Refer to appendix 2
Analogue proportional
Stylus length 100 mm (3.937 in)
4.8 V/mm
5.1 V/mm
5 V/mm
Stylus length 150 mm (5.9 in)
4.1 V/mm
4.3 V/mm
5.3 V/mm
16