Functions
7.1 Output settings
Digital output
Digital output can be used to configure various settings:
● Pulse (volume/pulse, pulse output, pulse width, pulse polarity, and time constant), see
● Frequency (frequency output, max frequency, and time constant), see frequency menu
● Error settings (level and number), see error level menu diagram and error number menu
● Limit settings (number of setpoints, setpoint settings, and hysteresis), see direction/limit
● Batch settings (quantity, time and counter settings, and time constant), see batch menu
Relay outputs
Relay output can be used to configure various settings:
● Error settings (level and number), see error level menu diagram and error number menu
● Limit settings (number of setpoints, setpoint settings, and hysteresis), see direction/limit
● Batch settings (quantity, time and counter settings, and time constant), see batch menu
● Cleaning (cycle time), see cleaning menu diagram.
48
pulse menu diagram.
diagram.
diagram.
menu diagram.
diagram.
Note
Batch settings
Only MAG 6000.
Not available in MAG 5000, MAG 5000 CT and MAG 6000 CT.
Note
When relay is set to batch function, pulse/frequency is not available on digital output.
diagram.
menu diagram.
diagram.
Note
Batch settings
Only MAG 6000.
Not available in MAG 5000, MAG 5000 CT and MAG 6000 CT.
Note
Cleaning
If a cleaning unit is installed together with transmitter, relay output must always be used
to operate this unit. It cannot be used for other purposes.
SITRANS F M MAG 5000/6000
Operating Instructions, 12/2013, A5E02338368-002